Reference
103
Sprint when to begin a test from a PLC through the
tester’s I/O port. Sprint can also send quantified test
results over its RS-232 output. See I/O
Pressure Decay
: Also called pressure drop or delta pressure.
The pressure change from the first measurement to the last
measurement made during the test phase of a test. See
Reject Level
Pressure Decay Test
: Pressure decay testing is used to test
products for leaks by trapping pressure inside a product
and then measuring pressure loss. The abbreviation P
Decay is often used in this guide to refer to pressure
decay.
Pressure Error
: Pressure setpoints above and below test
pressure. One pressure error (a plus and minus value) is
used by Sprint to insure that a valid test is performed on
each product. A lack of test pressure that causes a drop in
pressure to exceed the negative pressure error value during
the fill period of product testing causes Sprint to stop
testing and display the error in the status box. If test
pressure is accidentally increased, Sprint also shows the
error in the status box on the run display.
Pressure Regulator
: The Sprint iQ can be configured to
support up to 3 regulator inputs depending upon the model
type and option configurations. (Also see “Electronic
Programmable Regulator” on page 111.)
The regulator inputs are defined as:
Main
: Default input which is used as the standard
common regulated supply for all test channels on a
sequential tester and common regulator concurrent tester
or channel 1 for the independent regulator concurrent
tester. The regulated supply can be a manual or electronic
pressure regulator, a vacuum regulator supplied by an