Quick Start
17
Setting Test Time
As test times are developed,
keep in mind the size and
material of parts to be
tested. Parts that are both
small and rigid typically
require short test times.
Parts that are both large and
flexible generally require
longer test times.
Avoid the Tendency to
“Over-Test”
Optimum setup values are typically established after trial and
error when working with known good and bad product samples.
One point to keep in mind is not to over-test product. Avoid
setting leak tolerances too tight with the idea that absolutely no
leak is acceptable. Through experimentation you’ll probably find
that a small pressure drop occurs simply due to how your product
behaves during pressurization. Many products create conditions
that merely appear to be small leaks when pressurized.