UIT2400 INSTRUMENTATION MANUAL REV A, May 2020 

6 OF 18 

To take calibrated light level readings, press display units until light level is selected. 

Note: 

If light level is selected prior to selecting a calibration/sensitivity factor, an error 

message Set Factor will appear in place of light level readings (see 15P1 for selecting 

factors). If OD and % Trans, are selected and no baseline 100% value has been set 

and error Set 100% will appear see 3G8 below to learn more about setting 100%) 

 

3G3. Integrate: 

(Requires Display Units set to light level) Summation of data, records 

exposure/dosage (ie. W/cm2 becomes J/cm2 in integrate mode). 

 

3G4. Min-Max:

 Holds the display at the present value until a new minimum (lower reading) 

or new maximum (higher reading) is measured, then updates the display accordingly. 

3G4a. Note:

  Min-Max also serves as a Peak FW. See section 15F. Min/Max (and FW 

Peak) 

 

3G5. Record: 

Starts and stops (toggles between) continuous recording of readings and 

creates a new file with time, instantaneous, min, max and average measurements.  

 

3G6. Zero:

 Takes a measurement and creates a zero value by subtracting the zero 

measurement reading from all future readings. (see 3F2 above to learn more about 

assuring an accurate zero) 

 

3G7. Hold

: Freezes the numerical display value on screen. 

 

3G8. 100%:

 Takes a measurement and saves it as the 100% baseline for transmission and 

O.D. 

 

3G9. Capture:

 Takes a single measurement. Captured measurements are appended to 

the single sample file. 

 

3G10. Quick Start Care and Handling:

 

  Always store the meter in the carrying case when not in use and only ship in a 

carrying case with protective packing to prevent shock and damage. No items 

should be place above or under the meter inside the carrying case. 

  Do not expose meter to direct sunlight or high heat sources as the internal 

temperature can rapidly increase and cause erroneous readings and damage.  

  Standard operating temperature is rated at 0-40

0

 C. 

  Do not apply any force to the display.   

  Do not apply pressure to extended stand. 

  Opening the housing voids warranty, Housing is not water tight, avoid moisture, 

liquids and chemicals. 

Contact Ushio customer service for repair and calibration prior to shipment and obtain an 

RMA: [email protected] 

 

4. 

Internal Battery Pack

:

 

4A. About:

 Inside the UIT2400 is a 3.7V 5500mAh rechargeable lithium-polymer battery pack. 

A new fully charged battery pack typically last for 4-8 hours of continuous use (depending on 

display brightness settings and other feature settings).  

 

 

 

Содержание UIT2400

Страница 1: ...USHIO AMERICA INC 5440 Cerritos Avenue Cypress CA 90630 Tel 714 236 8600 Fax 714 229 3180 www ushio com UIT2400 Operation Manual Ushio America Inc...

Страница 2: ...Wiring 7 6 USB Ports 7 7 Display 7 8 Current Measurement and Limitations 7 9 Sample Time and Limitations 8 10 5V Bias for silicon sensors 8 11 Home Screen 8 12 Display Style Selection 8 13 Display Co...

Страница 3: ...ata retrieval C SED220 custom 222nm detector NIST traceable calibrated at 222nm D 8GB flash memory for measurement recordings E CC200 Case Protective carrying shipping case F Electrical Calibration Ce...

Страница 4: ...charging 3D Top Contains the sensor mount a 15 Pin D Sub connector and posts for securing the sensor to the device To connect the sensor align the sensor s male 15 pin connector to the meter s female...

Страница 5: ...and a scroll menu will appear select a new channel until you find the correct sensitivity factor as noted on your certificate If you do not locate your calibration factor and units from the certificat...

Страница 6: ...ure readings see 3F2 above to learn more about assuring an accurate zero 3G7 Hold Freezes the numerical display value on screen 3G8 100 Takes a measurement and saves it as the 100 baseline for transmi...

Страница 7: ...n the side of the UIT2400 device UPDATE Is used for code updates and requires the use of a memory stick not included DATA Accesses the internal 8GB memory device where data is saved and to charge the...

Страница 8: ...test settings Note There are multiple factors that affect the sampling time including the sensors rise time the light level software selection software settings and the method used to store and transf...

Страница 9: ...cases the level can be used as a go no go level indicating the risk of product failures when intensities are too low or too high YELLOW is typically used as a warning stage indicating the range has c...

Страница 10: ...ver time When in integrate mode the units of the readings change to include seconds For example W cm2 becomes W cm2 seconds or J cm2 W m2 becomes J m2 14D Hold Freezes the value on the numerical displ...

Страница 11: ...urrent 1uA 0 5sec is used else if current 1uA 1 0sec FAST samples at 20ms 50Hz Note Using trend at fast speeds taxes the CPU with processing both measurements numerical updates and graph updates This...

Страница 12: ...Interval Ranges from 1 second to 1 hour or sample for faster speeds Sample when selected the record interval and sample time are synched Ie if sample time is set 1 the meter will record a value every...

Страница 13: ...vice turns on 15M Use Scientific Notation ON OFF Scientific notation ON uses an exponent to show the decimal place when off a prefix is used Examples On 1 35 e 3 W cm2 Off 1 35 mW cm2 On 4 78e 6 W cm2...

Страница 14: ...calibration certificate 2 09e 09 see red arrow on cal cert on page 14 15Q 5 Resulting Light Level Allows the user to enter a known value for the light level under test and create a calibration factor...

Страница 15: ...ouble click on the Removable Disk within Devices with Removable Storage To open in excel you will need to select tab or space delimited 16D Deleting Stored Readings To delete files access the files th...

Страница 16: ...elect Location box 9 To save reading s touch the capture button Readings will be saved to the Survey_data_file csv file accessible fie the usb micro port 20 Info Lists product and software settings in...

Страница 17: ...f the meter Always store the meter in the carrying case when not in use and always ship in the hard plastic carrying case with protective packing to prevent shock and damage No items should be place a...

Страница 18: ...America customer service and obtain an RMA customerservice ushio com 24 Warranty The equipment you have purchased from Ushio America Inc has been expertly designed and was carefully tested and inspect...

Отзывы: