background image

UIT2400 INSTRUMENTATION MANUAL REV A, May 2020 

11 OF 18 

 

15 Settings:

 

The settings icons open the settings display which allow user customization.  

Settings can also be access by clicking on the setting drop down 

menu in the display screens. 

 

15A. Display Brightness

: Brightness selection ranges from 1-Dim with lowest battery 

consumption, to 10-Brightest with highest battery consumption. 

 

Radiometer Settings

 

15B Calibration factors:

 

15B1. Meter Factor:  

Selects calibration factors from the pre-programmed/available list 

ranging from 1 – 20. Factor selection allows readings in calibrated units such as W, 

W/cm2, Lux etc. 

 

15B2. Smart Detector Factor: 

 Smart detectors can also store up to 20 calibration 

factors. In order to use, or create cal factors that are saved in the sensor connector, Use 

Smart Detector Factor must be enabled, in 15B3. 

 

15B3. Smart Detector Factor ON/OFF:  

ON enable the use of the sensors cal factors. 

Off only allows customer to use factors save in the UIT2400 meter.  

 

15C. Sample Time:

 Allows the user to program how fast the meter takes a reading ranging 

from 100 milliseconds to 5 seconds with the default set to Auto. 

Auto will review the incoming signals intensity and apply the best settings. (faster with higher 

light levels, slower for smaller signals for better averaging)As a general rule of thumb, If current 

>= 1uA, 0.5sec is used, else if current <1uA, 1.0sec 

FAST samples at 20ms (50Hz) 

 

Note:  Using trend at fast speeds taxes the CPU with processing both measurements, 

numerical updates and graph updates.  This can be seen as a slight flicker on the display.  If 

this problem occurs, use meter (no graph) or connect to the computer and use Datalight III 

Meter or Flash. 

 

15D. Auto Gain Range:

 On or off. The UIT2400 has 3 relays that it cycles through when auto 

ranging from the lowest signals in the Pico amp range to the higher signals in the micro amp to 

1 milli amp range. Auto allows the software to control when the device will change from one 

relay to the next to allow the most accurate results based on incoming current levels. Auto gain 

range- off requires the user to set the range they will stay in during measurement.(see 15E 

below for information on setting the range.)    

 

15E. Gain Range: 

 When Auto Gain Range is set to off in section D above, Gain Range 

becomes accessible.  The user can select the range based the max current to be measured. 

Ie; 1 mA, 3uA or 300 nA.  Each relay allows the UIT2400 to measure approximately 3-4 

decades with the max being the saturation point and noise become problematic as the reading 

approaches .0005 from the max. 

 

To determine the range of current that applies to a specific light level reading, the user can 

select current in the display units (see section 14A for instructions)

 

or multiply a previous 

reading by the calibration factor used to obtain the reading. Ie. 1mW/cm2 (1e-3 mW/cm2) with 

Содержание UIT2400

Страница 1: ...USHIO AMERICA INC 5440 Cerritos Avenue Cypress CA 90630 Tel 714 236 8600 Fax 714 229 3180 www ushio com UIT2400 Operation Manual Ushio America Inc...

Страница 2: ...Wiring 7 6 USB Ports 7 7 Display 7 8 Current Measurement and Limitations 7 9 Sample Time and Limitations 8 10 5V Bias for silicon sensors 8 11 Home Screen 8 12 Display Style Selection 8 13 Display Co...

Страница 3: ...ata retrieval C SED220 custom 222nm detector NIST traceable calibrated at 222nm D 8GB flash memory for measurement recordings E CC200 Case Protective carrying shipping case F Electrical Calibration Ce...

Страница 4: ...charging 3D Top Contains the sensor mount a 15 Pin D Sub connector and posts for securing the sensor to the device To connect the sensor align the sensor s male 15 pin connector to the meter s female...

Страница 5: ...and a scroll menu will appear select a new channel until you find the correct sensitivity factor as noted on your certificate If you do not locate your calibration factor and units from the certificat...

Страница 6: ...ure readings see 3F2 above to learn more about assuring an accurate zero 3G7 Hold Freezes the numerical display value on screen 3G8 100 Takes a measurement and saves it as the 100 baseline for transmi...

Страница 7: ...n the side of the UIT2400 device UPDATE Is used for code updates and requires the use of a memory stick not included DATA Accesses the internal 8GB memory device where data is saved and to charge the...

Страница 8: ...test settings Note There are multiple factors that affect the sampling time including the sensors rise time the light level software selection software settings and the method used to store and transf...

Страница 9: ...cases the level can be used as a go no go level indicating the risk of product failures when intensities are too low or too high YELLOW is typically used as a warning stage indicating the range has c...

Страница 10: ...ver time When in integrate mode the units of the readings change to include seconds For example W cm2 becomes W cm2 seconds or J cm2 W m2 becomes J m2 14D Hold Freezes the value on the numerical displ...

Страница 11: ...urrent 1uA 0 5sec is used else if current 1uA 1 0sec FAST samples at 20ms 50Hz Note Using trend at fast speeds taxes the CPU with processing both measurements numerical updates and graph updates This...

Страница 12: ...Interval Ranges from 1 second to 1 hour or sample for faster speeds Sample when selected the record interval and sample time are synched Ie if sample time is set 1 the meter will record a value every...

Страница 13: ...vice turns on 15M Use Scientific Notation ON OFF Scientific notation ON uses an exponent to show the decimal place when off a prefix is used Examples On 1 35 e 3 W cm2 Off 1 35 mW cm2 On 4 78e 6 W cm2...

Страница 14: ...calibration certificate 2 09e 09 see red arrow on cal cert on page 14 15Q 5 Resulting Light Level Allows the user to enter a known value for the light level under test and create a calibration factor...

Страница 15: ...ouble click on the Removable Disk within Devices with Removable Storage To open in excel you will need to select tab or space delimited 16D Deleting Stored Readings To delete files access the files th...

Страница 16: ...elect Location box 9 To save reading s touch the capture button Readings will be saved to the Survey_data_file csv file accessible fie the usb micro port 20 Info Lists product and software settings in...

Страница 17: ...f the meter Always store the meter in the carrying case when not in use and always ship in the hard plastic carrying case with protective packing to prevent shock and damage No items should be place a...

Страница 18: ...America customer service and obtain an RMA customerservice ushio com 24 Warranty The equipment you have purchased from Ushio America Inc has been expertly designed and was carefully tested and inspect...

Отзывы: