USA57 Revision
1.2
73
Command:
Q
Status
Request
Type:
System Status Command
Syntax:
Q
Range:
0 to 3
0:
No
error
1:
Initialization
error
2:
Invalid
Command
3: Operand out of range
Description:
Query the Accuriss for its current status.
Returns the Ready/Busy status as well as any error conditions in the “Status”
byte of the return string. The return string consists of the start character (/), the
master address (0) and the status byte. Bit 5 of the status byte is set when the
Accuriss is ready to accept commands. It is cleared when the Accuriss is busy.
The least significant four bits of the status byte contain the completion code.
Errors in OpCode will be returned immediately, when errors in Operand range
will be returned only when the next command is issued. See Appendix B
Example:
/1
Q
<CR>
“/”
Is the start character. Lets Accuriss know that a command string will follow.
“1”
Is the device address, (this is preset at the factory).
“Q”
Requests Accuriss status
“<CR>” Tells Accuriss that the command string is complete.