9.0 Deleting stored Measurement Data / 

Display of All Memory Entries

Possibility to delete data by using key „Clear" (21).
Furthermore, it is possible to display the total num-
ber of measurement value entries on the display:

Display of Total Number of Measurement Value
Entries:

Set function switch (17) to one of the 5 measu-
rement ranges

Briefly press the key „Clear" (21). A short beep
signal is audible

The total number of all measurement values
saved is displayed

The display disappears after 2 seconds

Delete the Last Memorised Measurement Value:

Set function switch (17) to one of the 5 func-
tions

Press  key „Clear" (21) for approx. 2 seconds.

A short beep signal is audible

After 2 seconds, a further beep signal is audible
and the number of the remaining stored data is
displayed

Get down from the key “Clear” (21)

Deleting all Stored Measurement Values:

Set function switch (17) to one of the 5 func-
tions

Press  key „Clear" (21) for approx. 5 seconds.

Three short beep signals are audible.

„0" appears on the display. The memory is com-
pletely deleted, the „M" (11) symbol in the dis-
play disappears.

10.0 Energy Management 

Approximately 5 minutes after the last key operati-
on, the instrument switches off automatically (auto-
power-off). To switch the instrument on again, turn
the rotary switch from the „OFF" position to the de-
sired function. If the Battery is almost dead, the
symbol (8) appears in the indication. The piece of
equipment works furthermore, however without
claim on retention of the spezifications. If the bat-
tery discharges further, the piece of equipment
switches itselfs off when reaching the minimal bat-
tery tension. At empty battery cannot switch the
piece of equipment on.

11.0 Appropriate Usage 

The instrument may only be used under those con-
ditions and for those purposes for which it was built.
For this reason, in particular the safety references
(chapter 3.0),as well as the technical data including
environmental conditions (chapter 15.0) and the
usage in dry environments must be followed at all
times.

When modifying or changing the instrument,
the operational safety can no longer be gua-
ranteed.

If the battery case is open, the instrument must
not be connected to the measurements circuits

12.0 Maintenance

When using the instrument in compliance with the
instruction manual, no special maintenance is re-
quired.

12.1 Cleaning

If the instrument is dirty after daily usage, it is advi-
sed to clean it by using a damp cloth and a mild hou-
sehold detergent. Prior to cleaning, ensure that the
instrument is switched off and disconnected from
the external voltage supply and any other instru-
ments connected (such as EUT, control instru-
ments, etc.).

Never use acid detergents or dissolvants for
cleaning.

14

Deleting stored Measurement Data / Maintenance

Содержание TELARIS LOOP

Страница 1: ...TELARIS LOOP UNITEST Instruction Manual Cat No 8978...

Страница 2: ...2...

Страница 3: ...ement 8 6 2 Internal Mains Resistance Measurement 9 6 3 RPE RN RL Measurement 10 6 4 AC Voltage Measurement VAC 11 6 5 Frequency Measurement f 11 6 6 Contact Electrode 11 7 0 Display Indications 12 8...

Страница 4: ...ST TELARIS LOOP is charac terised by the following features A unit designed to allow the user to quickly esta blish earth loop impedance mains internal resi stance and prospective short circuit curren...

Страница 5: ...or all tasks In order to avoid electrical shock the valid sa fety and VDE regulations regarding excessive contact voltages must receive the utmost atten tion when working with voltages exceeding 120V...

Страница 6: ...s well as for trouble shooting Simultaneously to measurement the instrument TELARIS LOOP determines by calculation the short circuitcurrent Theshort circuitcurrentmust be known to guarantee that an ex...

Страница 7: ...for mains lead Test lead 14 LCD 15 Key Vc UL 50V 25V 25V for medical and agricul tural uses 16 Contact electrode 17 Measurement function switch off ZL IPSC Loop impedance short circuit current RI IPSC...

Страница 8: ...e mains lead to the mains connector 13 Select the measurement function R via the fun ction switch 17 Connect the test cables mains plug to UUT socket by respecting correct polarity Set a voltage conta...

Страница 9: ...ement results switch off all loads or disconnect them from the mains Connect the mains lead to the mains connector 13 Select measurement function via the function switch 17 Select contact voltage limi...

Страница 10: ...ressing the key Display 23 you may now select between the different measure ment variables If a value exceeds the measurement limit the Attention symbol appears If this result is sel ected with the ke...

Страница 11: ...t the measure ment instrument from UUT Also refer to chapter 7 0 No possibility to store the measurement re sults 6 5 Frequency Measurement Never apply tvoltages exceeding 300V to input sockets Connec...

Страница 12: ...ceeded Contact voltage limit 50 V Resistance measurement Zl Ipsc ZL 2 88 Contact voltage limit 50 V Resistance measurement Zl Ipsc RL excessive 19 99 Measurement is interrupted Contact voltage limit 5...

Страница 13: ...surement result by pressing key Store 19 Upon completion of storing a short signal is au dible and the symbol M 11 appears in the display for memory location In case of a full memory the symbol M blin...

Страница 14: ...rom the OFF position to the de sired function If the Battery is almost dead the symbol 8 appears in the indication The piece of equipment works furthermore however without claim on retention of the sp...

Страница 15: ...tery cells Never throw a battery into a fire as this could lead to an explosion Never expose batteries to humi dity Please consider your environment when you dis pose of your used batteries They belon...

Страница 16: ...01 Ohm 20 199 9 Ohm 0 1 Ohm 5 rdg 5 Digit 200 Ohm 1 999 kOhm 1 Ohm Internal Resistance RI Range Resolution Tolerance 0 05 19 99 Ohm 0 01 Ohm 5 rdg 5 Digit Resistance Neutral Conductor RN Range Resolu...

Страница 17: ...z 0 1 Hz 1 rdg 2 Digit 400V RI 1 M Contact voltage detection UL 25V 25V 5 15 UL 50V 50V 5 15 For large mains voltage fluctuations during measurements the measurement error may increase Auto power off...

Страница 18: ...18...

Страница 19: ...oumis un contr le technique permanent Ces m mes quipements de contr le sont calibr s r guli rement l aide d appareils de r f rence calibr s selon les di rectives et normes en vigueur dans les laborato...

Страница 20: ...e with unda maged warranty label Any damage due to dropping or incorrect handling are not covered by the war ranty If the instrument shows failure following ex piration of warranty our service depart...

Отзывы: