Carrying out Measurements
6.4 AC Voltage Measurement
Never apply voltages exceeding 300V to input
sockets.
Connect mains lead (13) to the mains connec-
tor
Select function VAC via the function switch (17)
Select the contact voltage by means of key (15)
in display
Connect the the test leads /mains plug to UUT
Read the measurement result on the display
If the voltage exceeds 300V, the message "At-
tention (1) + Limit (2) " is displayed. In this in-
stance, immediately disconnect the measure-
ment instrument from UUT.
Also refer to chapter 7.0.
No possibility to store the measurement re-
sults.
6.5 Frequency Measurement
Never apply tvoltages exceeding 300V to input
sockets.
Connect the mains lead to the mains connec-
tor (13)
Select the function f via the function switch (17)
Connect the the test leads /mains plug to UUT
Read the measurement result on the display
Also refer to chapter 7.0.
No possibility to store the measurement re-
sults.
6.6 Contact Electrode
Measurement of the present contact voltage is car-
ried out after setting the contact voltage limit. By
touching the contact electrode, exceeding the limit
is automatically indicated (>25V or 50V).