8.0 Display Indications / Error Messages
Measurement Range Message
Cause
UAC
>440V
overflow, voltage present exceeds
50Hz
440V AC
UAC
25V
frequency of AC voltage outside
---
range 10.0 ... 99,9Hz
UDC
>440V
overflow, voltage present exceeds
440V DC
UDC
>-440V
overflow, voltage present falls
500V
below -440V DC
RINS
>199,9M
Ω
insulation resistance exceeds or is
500V
greater than 199,99M
Ω
ZL, IPSC
>1999
Ω
loop impedance higher
than 1999
Ω
LOW-
Ω
>19,99
Ω
Low ohm resistance higher
than 19.99
Ω
LOW-
Ω
COMP
Test lead resistance has been
properly compensated.
RINS, LOW-W
Uext
External voltage present, measure-
ment may not be carried out.
RCD, ZL, IPSC
PE Contact Electrode
PE Error
is touched
RCD
Measurement of trip current, or
,
test of time delayed RCD.
RCD
Limit
Trip time exceeds 500ms or RCD
>500ms
has not tripped.
5V UL:50V
RCD
Limit
Trip time exceeds 300ms or RCD
300ms
has not tripped.
5V UL:50V
RCD
Limit
RCD has tripped at 345 ms. As trip
345ms
time exceeds 300ms, the RCD
5V UL:50V
was failled!
46
Display Indications / Error Messages
Содержание Telaris 0100