Chapter
9
16 (0x10) Write multiple registers
On slave holding registers, data is changed by the specified number from the specified address.
If broadcast (0) is specified, all slave holding registers of the same address are rewritten.
To request, specify the register address and the number of registers and data to change.
The slave transmits the contents of one register as expanded in double bytes.
* N = Number of registers
Example 1) Change the weighing function 2 (address 40047) to 111 (0x006F) and digital
low pass filter to 5Hz (*) in the 16-bit wide integer data area.
* Select the F701-P digital low pass filter setting from:
0: 1.5Hz, 1: 2Hz, 2: 2.5Hz, 3: 3Hz, 4: 4Hz, 5: 5Hz.
Write "5" for 5Hz in the example.
Keep in mind that the relative address for writing in weighing function 2 is 0x2E.
[Request]
Function
1
byte
0x10
Start address
2
byte
s
0x0000 - 0xFFFF
Number of registers
2
byte
s
0x0001 - 0x0078 (120)
Number of bytes
1
byte
2×N
*
Change data
N×2
byte
s
[Response]
Function
1
byte
0x10
Start address
2
byte
s
0x0000 - 0xFFFF
Number of registers
2
byte
s
0x0001 - 0x007B (123)
[Error response]
Error code
1
byte
0x90 (fu 0x80)
Exception code
1
byte
01 or 02 or 03
[Request]
Function
10
[Response]
Function
10
Start address; higher
00
Start address; higher
00
Start address; lower
2E
Start address; lower
2E
Register; higher
00
Register; higher
00
Register; lower
02
Register; lower
02
Number of data bytes
04
Weighing function 2; higher
00
Weighing function 2; lower
6F
Digital low pass filter; higher
00
Digital low pass filter; lower
05