104
11
SUPPLEMENTS
104
SU
PP
LE
M
EN
TS
Chapter
11
(Calibration error)
The load cell output exceeds the span adjustment range of the F701-P. Check to see if the load cell
is properly loaded and if the rated output value of the load cell is within the span adjustment range,
and redo span calibration.
(Calibration error)
Calibration is not completed properly because the indicated value of the F701-P fluctuates during
calibration.
Adjust the stable setting parameters (period and range), check that "
" lights, and then redo
calibration.
■
Checksum error display
(Checksum error)
This error is displayed if a checksum error is given during self-check.
■
F-RAM check error display
(F-RAM check error)
This error is displayed if an F-RAM check error is given during self-check.
* Checksum error display or F-RAM check error display indicates a failure.
Ask us or your distributor for repair.
The span adjustment range of the F701-P is 0.3 to 3.0 (mV/V).
Since display up to the capacity is ensured during span calibration, calibration
fails if the load cell output is under 0.3 (mV/V) (
) or over 3.0 (mV/V)
(
) when the capacity is reached.
Capacity
0
0.3
3.0
Load
Load cell output
[mV/V]
Span adjustment range
Point
Содержание F701-P
Страница 1: ...01AUG2020REV 1 07 WEIGHING INDICATOR F701 P OPERATION MANUAL...
Страница 11: ...Contents X Contents X M E M O...
Страница 116: ......