Chapter
9
- Transmission on format 3
* If two or more F701-P are connected, do not specify continuous mode.
■
Modbus-RTU
■
Transmission delay time (setting mode 5-4)
Set this when the master equipment cannot process responses from the F701-P.
*1
O : Overload (LOAD, OFL)
S : Stable
M : Unstable
H : Hold
Order of priority: H > O > (S or M)
*4
N : Near zero OFF
Z : Near zero ON
*2
A : Zero tracking OFF
T : Zero tracking ON
Z : Zero alarm (ZALM)
Order of priority: Z > (A or T)
*5
Unit display
0 : None
3: kg
1 : t
4: lb
2 : g
*3
Upper/lower limit comparison and
over/under comparison modes
H : Upper limit/over ON
L : Lower limit/under ON
G : Upper limit/over and
Lower limit/under OFF
N : Upper limit and lower limit ON
F : Comparison OFF
Priority : N, F
>
(H or L)
>
G
Discharging control Mode
0 : SP1, SP2, SP3, Complete OFF
1 : SP1 ON
2 : SP2 ON
3 : SP3 ON
C : Complete ON
F : Comparison OFF
Priority : C, F
>
3
>
2
>
1
>
0
*6
Decimal place
0 : None
1 : 0.0
2 : 0.00
3 : 0.000
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
W T
*1 *2 *3 *4
±
HEADER
5-digit net weight
Sign (gross weight)
+ or -
5-digit gross weight
*5
SOH
STX
ETX BCC
Sign (net weight)
+ or -
±
BCC calculation target
Each ASCII code of SOH, STX, and ETX
SOH
:
01
STX
:
02
ETX
:
03
BCC calculation method
Each code targeted for BCC calculation is
expressed in hexadecimal notation, and the
result of XOR with all data enters BCC.
*6
22 23 24 25
I D
Transmission delay time
Input range/0 - 99
3.5 char + transmission delay time (msec)
Frame
}
Frame
}