3
1
OUTLINE
3
O
U
TL
IN
E
Chapter
1
(4) Final
(5) None
Accumulation count
An accumulated count is displayed.
No more accumulations are made after "9999", which cannot be
counted up any more. It is initialized to "0" by performing
Accumulation Clear.
Accumulation value
An accumulated value is displayed.
It is initialized to "0" by performing Accumulation Clear.
Final setting value
"
Final (target value)" set by setting mode 0-9 is displayed.
③
Unit display
A selection can be made from t, g, kg, lb, and none by setting.
④
Status display
Lights when the weight value
≦
near zero setting value.
* The type of weight value is determined by near zero comparison mode
(Weighing function 1 in setting mode 1-4).
Lights at the true zero point (0±1/4 scale division)
(when 1/4 scale division in setting mode 4-5 "display selection" is off).
Lights at the true zero point (0±1/4 scale division) and the central point of the scale
interval of the indicated value {
indicated value±(1/4×min scale division)
}
(when 1/4 scale division in setting mode 4-5 "display selection" is on).
Lights when the zero tracking function is working.
Lights when the weight value is held.
Blinks when the total amount of digital zero exceeds the DZ regulation value by
digital zero or zero tracking operation.
Lights when the weight value is stable.
* For the definition of stability, see
"5-6.Motion detection(MD) (setting mode 1-5,
Lights when the LOCK(soft) is on.
Final value