45.
Measuring that resolution and BPP match the reference frame
Testing the CRC of the incoming video against the reference frame’s CRC values captured
and defined earlier. The number of mismatches found during the are shown.
Test Results
Finally, it shown if the test passed. In the UCD Console GUI it is set how many
mismatched on the test are allowed. In this case 0 mismatched were found and the test
passed. The automated CRC based single frame video stability test took 2.14 seconds to run