Manual – Starflow QSD Model 6527B and 6537A
Unidata Manual - 6527B and 6537A Starflow QSD User Manual 08 05 2023.docx
Page 40
Pipe Diameter
Clamp
Bands
Joiners
Up to 600mm
6705A
6705D
-
600 - 1200mm
6705B
2 x 6705D
1
1200 - 1800mm 6705B
3 x 6705D
2
Note
: Pipes less than 150mm in diameter are NOT RECOMMENDED for Starflow QSD
installation due to the instrument’s size. Do the following:
1. If the inside circumference of the pipe is ACCURATELY known, then cut the band
assembly to this length, LESS 25mm, for fitting clearance. Circumference (length) =
Diameter x 3.14
2. Arrange the clamp, band(s) and joiner(s) so that the Starflow QSD unit will be positioned
at the bottom (invert) of the pipe and an Expanding Clamp at the top (obvert) of the pipe
(see diagram).
3. Drill 4 x 5mm holes, 85mm, spaced in the centre of the band to locate the instrument.
Use the Starflow QSD mounting clamp 6527M as a guide.