5.1.3.2.
Pressure Depth Zero Offset
Example 1 Examine:
Send – 0X7! <CR><LF>
Response – 0+30<CR><LF> The current pressure depth offset setting is 30mm.
Example 2 Change:
Send – 0X7+100! <CR><LF>
Response - 0+100<CR><LF> Set pressure depth offset to 100mm.
Example 3 Change:
Send – 0X7-100! <CR><LF>
Response – 0-100<CR><LF> Set pressure depth offset to -100mm.
5.1.3.3.
Enable Velocity Direction
Example Change (negative and positive velocity direction):
Send – 0X5+26! <CR><LF>
Response – 0+0+4+5+7+8+16+17+18+19+20+21+22+24+26+27+30<CR><LF>
5.1.3.4.
Changing Velocity Range
2 = 3345mm/s
3 = 1673mm/s (Default)
4 = 836mm/s
5 = 418mm/s
Example 1 Examine:
Send – 0X9! <CR><LF>
Response – 0+3<CR><LF> (Range reported as 3 – 1673mm/Sec)
Example 2 Change:
Send – 0X9+5! <CR><LF>
Response - 0+5<CR><LF> (Change range to 5 – 418mm/Sec)