UTL8200 User Manua
25
Figure 5-1-2 Settable Status (White, Unlocked)
Setting steps:
1. Use the knob or direction buttons to position the cursor on the field [Language], as
shown in Figure 5-1-1. The text is yellow, showing the current setting status is locked
and cannot be modified.
2. Press Enter to unlock the current settings. At this time, the text is white, as shown in
Figure 5-1-2, indicating that it is in the settable status.
3. The users can view setting options with the pulse knob, and the text of each option is
still white.
4. Press Enter to confirm the selected option. The option text is yellow and the setting is
complete.
5. Press the ESC button to return to the previous level, and press the ESC button again
to enter the main operating interface.
The other setting operations are the same.
5.2 Parameter Setting
The parameter setting is used to set the operating parameters and protection parameters
of the system. The detailed setting range is related to the model.
Parameter setting
Setting range
Description
Run time
0~99999s
Set the time of each load. No matter in which
mode, the load will stop automatically when it
operates to the corresponding operating time.
Short time
0.1~99999ms
Set the time for a single short circuit test
V limit
0~150V
Set the voltage value of over-voltage protection
I limit
0~20A (40A)
Set the current value of over-current protection
P limit
0~200W (400W)
Set the power value of over-power protection
V on
0~150V
Set the load voltage at the beginning of each
operation
V off
0~150V
Set the low voltage to automatically end loading
V trigger
0~150V
In list mode, the load will operate automatically
when it detects a voltage value higher than the
self-start voltage at the measuring terminal.
OVP
ON/OFF
When the input terminal voltage is higher than
the protection voltage, directly short-circuit the
input terminal.