UTL8200 User Manua
22
of the screen (UTL8212 only).
2. Press Shift +Mode to enter the main menu interface.
3. Use the knob to select the parameter setting, and then press Enter to enter the
parameter setting interface.
4. Use the knob to set the short circuit time, and press Enter to complete the setting.
5. Press the Mode button to enter the mode selection interface, select the CC/CV/CP/
CR mode, press Shift+
→(short) to enter the short circuit mode, and the symbol [Short]
will appear. Set the corresponding current value with the knob and press the
ON1/ON2 button to complete the short circuit test.
4.3 Parameter Input and Operating Control
4.3.1 Parameter Input
There are two ways to input parameters: rotate the knob or press the direction buttons.
When the set parameter is out of range, it will not be saved by pressing Enter. Please
reset the correct parameter.
4.3.2 Operating Control
After the electronic load is powered on, it is in an unloaded status. Control the input switch
of the load by pressing the ON1/ON2 button on the front panel. When the ON button
indicator light is on, it means it is in the loaded status, and when the ON button indicator
light is off, it means it is in the unloaded status.
4.4 Alarm Prompts
1. Overvoltage protection: load input voltage is higher than the set value of overvoltage
protection.
2. Overcurrent protection: load current is higher than the set value of overcurrent
protection.
3. Power protection: the load power is higher than the set value of over power
protection.
4. Polarity error: The positive and negative poles of the load input terminal are reversed.
5. Undervoltage protection: When the load is operating, it is detected that the input
voltage is lower than the unloading voltage set value.
6. Data overrange: The data entered is out of range (the upper limit of parameters such
as voltage/current/power/time).
7. Position error: If the voltage position/current position is set to the low position, when
the voltage/current value exceeds the range of the low position during load operation,
the error information will be displayed. The users should return to the parameter