UTL8200 User Manua
20
When using the check function, the value range of the item cannot be checked in the
single constant value mode. For example: In CC mode, only the voltage value and power
value can be checked, not the upper and lower limits of the current. After setting the
parameters, press ON1/ON2 to transfer the list files to the corresponding channel.
Operation steps:
1. After starting up, first press the CH button to select the operating channel. The yellow
wireframe area is the selected channel. And CH1 or CH2 will be displayed at the top
of the screen (UTL8212 only).
2. Press the Mode button to enter the mode selection interface, select the list mode with
the pulse knob or direction buttons, and then press Enter to enter the list mode setting
interface.
3. Select the modified parameter with the pulse knob, press Enter to modify, and press
Enter again to save the modification.
When setting list parameters (group, step, repeat, mode), first press the Enter button
to enter the modifiable status, then use the knob to select the proper mode or adjust
to the appropriate value, and finally press the Enter button to complete the setting.
When setting mode parameters (mode, check), position the cursor on the parameter
that needs to be set, press Enter to view different parameter options, stop pressing
on the selected option, and then press the direction buttons or rotate the pulse knob
to enter the next parameter setting.
When setting the mode parameters (value, time, min, max), position the cursor on the
parameter that needs to be set and press Enter. At this time, the setting xxxx at the
bottom of the screen will change to a specific value which can be adjusted by rotating
the pulse knob, and then press Enter to complete the setting.
Note:
During the setting process, if the set value exceeds the limit, an ERROR
pop-up window will appear. Press the Enter button and the window will disappear.
The users need to adjust the value to the range with the pulse knob to complete the
setting; otherwise the pop-up window will continue to appear.
4. After setting the parameters, press the ON button to enter the list test interface
(channel 1: ON1, channel 2: ON2).
5. Press the ON button to start the operation, and press the ON button again to stop it.
4.1.7 Battery Test
Battery mode is used to detect the battery capacity. Battery capacity is an important
indicator which reflects battery life and reliability. The voltage will decrease as the
discharge time increases when testing the battery capacity, so the stop voltage should be
set. When the stop voltage is reached, the test ends.
Battery mode parameters:
Parameters
Description
Mode
Battery discharge mode: CC/CR/CP
Load value
Set loading value