Temperature Measurement
Issue V 1.0
(
2020-10-19
)
39
3.6 Measure Verify
NOTE
⚫
This operation does not need to be started, and only needs to be started when the
measuring device is provided by a third-party measurement corporate.
⚫
If the function is enable, the temperature measurement function of the human body is
paused, and only the temperature of the object (standard black body) in the ID area is
working.
Table 3-6
Parameter of measure verify
Parameter
Description
Setting
Enable
Enable to measure verify
[Setting method]
enable
Area ID
Choose ID to set area of
measuring object
[Setting method]
Select from drop-down
list .
Emission rate
Emission rate is the thermal
calibration device’s base
parameter, the general
blackbody’s is 0.98.
[Setting method]
Input value
Distance
Distance is the actual horizontal
distance between measuring
object and the camera.
[Setting method]
Input value
----End