SMART SCREEN DISPLAY
USER MANUAL — 74-36121-ENG R2
53
Patient risk management screen
The Premium interface displays all bed options. The selected bed exit control level of detection will be highlighted.
WARNING
• The bed exit detection system must be used with patients with minimum weight of 50 lb/22.6 kg and the
auto-arm function must be used with patient with minimum weight of 70 lb/31.8 kg. If not avoided, the bed
exit detection system may not detect all movements and the functionality of the detection system could be
affected.
ATTENTION
• A zeroing procedure must be performed each time new accessories are installed or removed from the bed,
or each time a new patient is admitted in the bed. Failure to perform the zeroing, or a zeroing that is not
performed correctly, may render the bed exit detection system non-functional and increase the risk of a
patient fall.
• Patient must always be out of the bed during the zeroing procedure.
Select:
•
Bed exit management screen
to set level of
detection (Detection 1, 2, 3), auto-arm function
and alarm settings.
•
Bed status management
to set the brakes,
siderails, backrest and/or elevation system
position for bed status monitoring.
•
View log
to access a log listing setting
activation or deactivation and to highlight bed
exit control activation and alarm.
•
Inform
to set up a screen saver configured
to inform caregivers with color coding and/or
different pictograms from Pictograms library
and/or siderail configuration requirements based on patient condition.
•
Zero
to “zero” the bed exit settings. Menu mentions that patient shall not be in bed. Press ZERO to
proceed, message box mentions not to touch bed, or press CANCEL to cancel command.
IMPORTANT
A simplified interface displaying less options is offered for a simplified use. It is possible to change the interface
through the Preference menu in the Maintenance menu. On the Standard interface, the Inform button is not
available. If PIN protected, contact maintenance for access.
Buttons from Standard interface operate the same way as the buttons from the Premium interface.
Patient Risk Management
November 1st I 12:00
Patient Risk Management
November 1st I 12:00
Bed Exit
View Log
Inform
Bed
status
Bed Exit
View Log
Inform
Bed
status
Zero
!
!