GETTING STARTED
26
USER MANUAL — 74-36121-ENG R2
Configuration steps for server connection settings (optional)
Bed equipped with the connectivity option must be properly configured and tested during the start up procedure
so the available information of the bed can be properly relayed to the web application.
ATTENTION
•
Configuration can be performed from the display screen but the settings are limited; use the Umano Connect
Configurator tool to perform a configuration of all advanced settings.
IMPORTANT
Only
Umano Medical
representatives are authorized to perform a system update with an approved software.
Umano Medical
takes no responsibility for malicious activity that may result from lack of current, appropriate IT
security practices.
The network administrator must follow the steps and enter the proper configuration information.
If performing the configuration from the display screen, the display screen interface keyboard allowing the settings
configuration is limited; if any special characters are required, it is recommended to use the Umano Connect
Configurator tool.
1.
Use the Umano Connect Configurator to perform the network configuration. Refer to the user guide for
procedure included directly in the configurator tool.
2. From the display screen, enter the Bed location menu through Preference menu to enter the location number
3. Verify that the connection pictogram on the display screen turns green to indicate that bed is connected.
4. Verify that the bed appears in the “connected beds” list in the connectivity web application.
5. Verify that the bed exit detection signals are communicated through the web application by generating a false
bed exit audible signal (e.g. arm the bed while empty).
6. If applicable, verify that the bed status management signals are communicated through the web application
by activating motion of the elevation system, braking system and siderails. Also take a reading of the scale
and compare with the values and messages received at the nurse station. All status of the bed must be
properly communicated to the web application.
Bed serial number
Functions
Proper communication?
Bed exit system signals
Backrest angle
Elevation system height
Siderails positioning (optional)
Brakes positioning (optional)
Scale values
!