5.1
Installation
31
5
5.1.3.2 Connection of Cooling Water Line
(1) Supply the coolant water to coolant water IN connection of the solenoid valve for the
coolant water as shown in Fig. 5-8, and drain the cooling water pipe from coolant water
OUT connection. The valve to isolate (Lockout / Tagout) cooling water should be installed
at user equipment.
(2) Connect cooling water pipes to these pumps. When connecting the cooling water pipes,
screw the pipe joint while also holding the nozzle of the pump with a spanner in order to
avoid deforming the cooling water pipe, as illustrated in Fig. 5-9.
(3) The water valve is closed and coolant water supply is stopped at AC input OFF. Customer
prepares the cross valve and piping, the pump is bypassed, and the coolant water can be
thrown. Addition and the pipework of the valve in the customer are necessary.
Fig.5-8
Connection of Cooling Water Line
Fig.5-9
Water Coolant Tubing Connections
PIPE JOINT
Содержание UTM2300 Series
Страница 2: ...This page is intentionally left blank...
Страница 10: ...viii Location of Label...
Страница 20: ...6 1 1 2 Descriptions This page is intentionally left blank...
Страница 21: ...2 2IDENTIFICATION AND FUNCTION 2 1 Pump Main Unit 2 2 Control Panel 2 3 External I F Panel...
Страница 25: ...3 3CONSTRUCTION AND PRINCIPLE 3 1 Pump Construction 3 2 Principle of Turbo Molecular Pumping 3 3 Controller...
Страница 30: ...16 3 3 3 Controller This page is intentionally left blank...
Страница 36: ...22 4 4 3 Standards Fulfilled This page is intentionally left blank...
Страница 50: ...36 5 5 5 Notes on Transportation This page is intentionally left blank...
Страница 72: ...58 6 6 8 Communication Specifications This page is intentionally left blank...
Страница 73: ...7 7GAS PURGE...
Страница 82: ...68 8 8 6 Turbo Molecular Pump Return Request This page is intentionally left blank...
Страница 95: ...10 10WARRANTY CLAUSES...
Страница 98: ...84 10 This page is intentionally left blank...
Страница 100: ...Index Index 2 This page is intentionally left blank...