4.1
Standard Specifications for Pump Unit
19
4
Instantaneous power interruption
(Note 6)
If power interruption is 1 second or less, pre-interruption
functions are maintained. If interruption exceeds 1 second, the
brake is activated, which allows the pump to be restarted after
resetting.
Water valve for dew interlock
This valve is closed during power supply OFF, and the coolant
water is stopped.
This valve is opened during power supply ON, and the coolant
water is supplied.
Alarm detection
Alarms
Pump temperature, Pump startup error, Overload and
Overspin for motor, Failure of magnetic bearing,
Control system malfunction (Overtemperature inside control
system or failure of drive circuit), Power interruption,
Dew condensation, etc.
Warnings
Failure of magnetic bearing, Control system malfunction
(Overtemperature inside control system),
Dew condensation (Water valve close.)
Protective functions
Alarms
STATUS orange lamp illuminates.
Power Interruption Alarm :
Decelerated while maintaining levitation using regenerative
power.
After decelerating to specified low speed, levitation is
stopped and rotor is supported by touchdown bearing. When
power is restored, the pump can be restarted by resetting.
Dew Alarm :
The pump is decelerated after dew condensation alarm is
detected.
When other alarms occur :
Stops operation or decelerates. Magnetic levitation is
continued.
Warnings
STATUS orange lamp flashes and pump continues to function.
(Dew Warning : Water valve is closed.)
Input electric
power
Voltage
Single phase 200 to 240 VAC ± 10 % (50 / 60 Hz ± 2 Hz)
Maximum power
1.2 kVA
Insulation
withstand
voltage
1500 V, 1 minute
Short circuit
current ratings
(SCCR)
200 A
Water (Note 7)
Flow rate
Pressure
Temperature
3 to 4 L/min
0.2 to 0.4 MPa
19 to 30 degrees C. (above dew point)
Water quality
Non-corrosive industrial water
Solid particle size (Max) : 0.025 mm
2
PH (@25 degrees C.) : 6.5-8.2
Electrical conductivity (@25 degrees C.) : 100-800 μS/cm
(Electrical resistivity : 1250-10000 Ω•cm)
Admissible ambient
magnetic field
Radial direction
Axial direction
3 mT
15 mT
Содержание UTM2300 Series
Страница 2: ...This page is intentionally left blank...
Страница 10: ...viii Location of Label...
Страница 20: ...6 1 1 2 Descriptions This page is intentionally left blank...
Страница 21: ...2 2IDENTIFICATION AND FUNCTION 2 1 Pump Main Unit 2 2 Control Panel 2 3 External I F Panel...
Страница 25: ...3 3CONSTRUCTION AND PRINCIPLE 3 1 Pump Construction 3 2 Principle of Turbo Molecular Pumping 3 3 Controller...
Страница 30: ...16 3 3 3 Controller This page is intentionally left blank...
Страница 36: ...22 4 4 3 Standards Fulfilled This page is intentionally left blank...
Страница 50: ...36 5 5 5 Notes on Transportation This page is intentionally left blank...
Страница 72: ...58 6 6 8 Communication Specifications This page is intentionally left blank...
Страница 73: ...7 7GAS PURGE...
Страница 82: ...68 8 8 6 Turbo Molecular Pump Return Request This page is intentionally left blank...
Страница 95: ...10 10WARRANTY CLAUSES...
Страница 98: ...84 10 This page is intentionally left blank...
Страница 100: ...Index Index 2 This page is intentionally left blank...