Installation
3-8
ULVAC CRYOGENICS INCORPORATED
2.
Connect the refrigerator power cable from the compressor to the cryopump.
3.
Connect the ground.
3.
Connect the connector side of the input power cable to the compressor.
Connect the other side of the input power cable to the power source.
Note : If customer-supplied input power cable is used, be sure that it is equivalent to
UCI-supplied cable.
Table 3-2
Input Power Cable Specifications
Item Symbol
Specifications
Input power cable
INPUT
POWER
Copper wire / 600VAC, 3 conductors with ground
(*1)
1.25mm
2
(*1)
The color of the ground wire is green.
CRYOPUMP
FLEXIBLE HOSE(SUPPLY)
FLEXIBLE HOSE(RETURN)
REMOTE/RESPONSE
INPUT POWER CABLE
COLD HEAD INPUT
POWER CABLE
IN
COOLING WATER
OUT
【
C10T
】
Figure 3-5
Compressor Installation
Содержание C10T
Страница 2: ......
Страница 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 14: ...S 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 16: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 32: ...Operation 4 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 36: ...Disconnection and Storage 5 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 44: ...Troubleshooting A 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 47: ...Compressor Unit Instruction Manual ULVAC CRYOGENICS INCORPORATED B 3 B 4 ...
Страница 48: ...Compressor Unit Instruction Manual ULVAC CRYOGENICS INCORPORATED B 5 B 6 ...
Страница 50: ...Flexible Hose C 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 54: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 56: ...This page intentionally left blank R 2 ULVAC CRYOGENICS INCORPORATED ...