44
NT UF 841 CP GB (d) / June 2015
Enter the minimum and maximum limits of the dimension associated with the flow graph. These limits mark
the boundaries of the vertical display of the flow graph.
The flow graph Period setting is used to define the time interval between two successive points. The
possible values are: 1 s, 5 s, 10 s, 30 s, 1 min, 2 mins, 5 mins, 10 mins, 15 mins, 30 mins, 1 hr, 2 hrs, 6 hrs,
12 hrs and 24 hrs.
Example: If the period of the flow graph is 1 min, each point of the curve represents the average value of
the magnitude associated with the flow graph over 1 min and the whole screen shows the evolution of the
flow over the last 144 minutes.
4) Chord settings
For each chord, the two settings to be selected are the reference number of the probes used and the probe
fitting type:
If the reference number of the probes you are using does not appear in
the list, please contact Ultraflux.
5) Type of fluid
In "simple" configuration mode, the fluid must be water at ambient temperature (cannot be modified).
6) Definition of the pipe
The pipe diameter, thickness and material must be defined.
Содержание Uf 841
Страница 4: ...4 NT UF 841 CP GB d June 2015 CHAPTER 1 OVERVIEW...
Страница 12: ...12 NT UF 841 CP GB d June 2015 CHAPTER 2 SAFETY INSTRUCTIONS...
Страница 17: ...17 NT UF 841 CP GB d June 2015 CHAPTER 3 INSTALLATION AND WIRING...
Страница 25: ...25 NT UF 841 CP GB d June 2015 CHAPTER 4 INSTALLING A MEASURING POINT...
Страница 27: ...27 NT UF 841 CP GB d June 2015...
Страница 35: ...35 NT UF 841 CP GB d June 2015 CHAPTER 5 USING AND CONFIGURING THE Uf 841...
Страница 64: ...64 NT UF 841 CP GB d June 2015 CHAPTER 6 FUNCTION ENGINE...
Страница 68: ...68 NT UF 841 CP GB d June 2015 CHAPTER 7 PC SOFTWARE...
Страница 82: ...82 NT UF 841 CP GB d June 2015 Appendix I Characteristic echo signals...
Страница 85: ...85 NT UF 841 CP GB d June 2015 Appendix II Speed of sound in water...