31
NT UF 841 CP GB (d) / June 2015
4.3.2 External probes (clamp-on)
External probes may be mounted in several configurations depending on the number of ultrasound wave
reflections on the wall of the pipe:
Direct or / configuration:
N (or Z) configuration:
Reflex or V configuration:
W configuration:
Note:
The longer the travel, the more accurate the measurement.
On the other hand, the ultrasound echo will be weaker the longer the distance and therefore difficult to
measure
.
A compromise must therefore be found between accuracy and ease of finding the ultrasounds.
This compromise depends on the application (fluid, quality of the wall, diameter, etc.).