89
NT UF 841 CP GB (d) / June 2015
•
The measurement differs from the predicted flow
Actions to take
o
Check that the probes are far enough away from hydraulic disturbances.
o
Check the precise dimensions of the section of the pipe or the channel at the location of the
measurement and the position of the probes.
o
Have faith in your Uf 841.
•
The message "relay overflow" appears permanently or intermittently in the case of a
relay output configured in totalizer mode
Cause:
o
Too many pulses output within the allocated time.
Actions to take
o
Adapt the pulse weight and/or the width based on the flow of your application (see the
"Totalizer" section 5.4.4 - point 5).
•
The message "open loop" appears permanently when setting a current output
Cause:
o
Break in the loop.
•
The message "outside limits" appears permanently or intermittently when setting an
analogue input
Cause:
o
Current or voltage outside the range of use.
Содержание Uf 841
Страница 4: ...4 NT UF 841 CP GB d June 2015 CHAPTER 1 OVERVIEW...
Страница 12: ...12 NT UF 841 CP GB d June 2015 CHAPTER 2 SAFETY INSTRUCTIONS...
Страница 17: ...17 NT UF 841 CP GB d June 2015 CHAPTER 3 INSTALLATION AND WIRING...
Страница 25: ...25 NT UF 841 CP GB d June 2015 CHAPTER 4 INSTALLING A MEASURING POINT...
Страница 27: ...27 NT UF 841 CP GB d June 2015...
Страница 35: ...35 NT UF 841 CP GB d June 2015 CHAPTER 5 USING AND CONFIGURING THE Uf 841...
Страница 64: ...64 NT UF 841 CP GB d June 2015 CHAPTER 6 FUNCTION ENGINE...
Страница 68: ...68 NT UF 841 CP GB d June 2015 CHAPTER 7 PC SOFTWARE...
Страница 82: ...82 NT UF 841 CP GB d June 2015 Appendix I Characteristic echo signals...
Страница 85: ...85 NT UF 841 CP GB d June 2015 Appendix II Speed of sound in water...