21
To protect the instrument, against being dropped accidentally, use the wrist strap.
WRIST STRAP
:
HEAD SET JACK
This is where you plug in the headset. Be sure to plug it in firmly until it clicks.
:
This Jack receives the plug from the recharger. The recharger is designed to plug into a standard electrical
receptacle.
RECHARGE JACK
:
Keep a backup battery fully charged using the Charging Pod. This is a battery recharge docking station for
charging Ultraprobe Batteries (Lithium Ion). This pod will charge the standard batteries that come with the
Ultraprobe 15,000 while removed from the metered pistol housing.
CHARGING POD:
USERS INSTRUCTIONS
1.
This module plugs into the front end of the instrument.
TRISONIC
TM
SCANNING MODULE
:
2.
Align the pins located at the rear of the module with the four jacks in the front end of the Metered
Pistol Housing (MPH) and plug in.
3.
For general use position the frequency selection to 40 kHz.
4.
Start to scan the test area.
The method of air borne detection is referred to as
"Gross to Fine"
. Start at maximum sensitivity (S=70),
constantly reducing the sensitivity and following the bar graph amplitude display to the loudest point. If
there is too much ultrasound in the area, reduce the sensitivity, place the RUBBER FOCUSING PROBE
(described below) over the scanning module and proceed. Every time the sound level rises to a point
where it is difficult to follow, reduce the sensitivity again and again until you are able to follow the test
sound to its' loudest point.
METHOD OF AIRBORNE DETECTION:
The DHC2-HH headphones are designed to be worn with hard hats. To use, simply plug the headset cord
into the headset Jack on the pistol housing, and place the headphones over your ears.
HEADSET:
To use, simply slip it over the front of the scanning module or the contact (stethoscope) module.
RUBBER FOCUSING PROBE
:
NOTE
removing the Rubber Focusing Probe.
: To prevent damage to the module plugs, always remove the module BEFORE attaching and/or
1.
This module plugs into the front end of the instrument.
LONG RANGE MODULE (LRM):
2.
Align the pins located at the rear of the module with the four jacks in the front end of the Metered
Pistol Housing (MPH) and plug in.
3.
For general use position the frequency selection to 40 kHz.
4.
Start to scan the test area.