19
CF INFO
This feature allows you to view file names stored on the Compact Flash Card. When a sound is recorded and a
WAV
file is
saved, two files are written to the Compact Flash Card: a text file which contains all stored record fields for the specific record
number associated with the
WAV
file and the
WAV
file. When CF info is selected, 5 selections will be viewed on the screen.
You may click the Sensitivity Dial to select/enter any of these.
1.
Free Space & Info (This will display the number of Bytes Free and the Remaining Recording time on the card).
2.
Display Files (The files stored on the CF card)
3.
Delete Files
To delete a file:
a. Click to enter, the File Number will blink.
b. Press
Sens
to
DELETE
or Spin to the next file (the file you wish to delete)
c. If you don’t want to delete a file, press
ENTER
to exit.
4.
Delete ALL Files
5.
EXIT to MAIN
WARNING:
Never remove or insert the Compact Flash Card with the instrument turned on.
TO STORE A READING:
1.
Continue to
“Click”
the
Sensitivity Control Dial
to move the indicator to the bottom of the meter until
PRESS ENTER
FOR
: is blinking. Then
Spin
the dial until
STORE
Record is displayed. press the
Yellow Enter Button
. This puts the
instrument in the Data Storage mode. In the data storage mode the
Display Panel
changes.
2.
The screen will display the following:
Save To Record
(the current record number is displayed, (for example
003)
(This
allows you to determine whether to store the information in this particular record location or to change the location.)
Turn
SENS.: Change
(If you want to change the store location this prompts you to turn (Spin) the
Sensitivity Control Dial
to
change the location). Press the
Sensitivity Control Dial to EXIT
(If you do not wish to store the information, Click the
Sensitivity Control Dial
to return to operation mode)
Click (Press) Enter
to Store (If you want to store the information,
Click
Enter
, a screen will pop up to confirm the information has been stored.
3.
To view previously stored information, Click the
Sensitivity Control Dial
to the bottom of the
Display Panel
until
“PRESS
ENTER FOR”
is blinking and spin to
Storage Display
. Press
Enter
to view the stored data. To view information in other
storage locations, click the
Sensitivity Control Dial
until the storage number blinks. The storage number in the
Storage
Display
mode is located in the upper left corner. Spin the
Sensitivity Control Dial
to the desired storage location.
4.
If you wish to view all stored information during inspections, you may set the screen to
Storage Display
.
Click the
Sensitivity Control Dial
until
“PRESS ENTER FOR”
is blinking. 2. If
Storage Display
is not shown, spin until
Storage Display
is shown and blinking. 3. Click the
Yellow Enter button
. The
Display Screen
will be split. All
inspection data will be displayed on the top half while the stored information will be displayed on the bottom half of the
Display Screen.
STORING ABCD INFORMATION
To Store information when the ABCD Function has been set:
1.
When Click
ENTER
for
Store (A)
is blinking, press
ENTER
2.
Then Click
ENTER
for
Store (B)
will then blink, press
ENTER
3.
Then Click
ENTER
for
Store (C)
will blink, press ENTER
4.
Then Click
ENTER
for
Store (D)
will blink, press
ENTER
5.
The function will then shift to
STORE DATA
6.
Press
ENTER
and all four
dB
levels will be stored at the selected Record Number.
FREEZING A READING ON THE DISPLAY PANEL
If in a test situation where it is difficult to view the display panel, when in the Timed Off Mode the decibel reading can be frozen for
viewing. If using the Stethoscope Module, continue to touch the test point and then release the trigger. Remove the instrument and
view the panel. Be sure to view within the set timed off period (ex: within 5 seconds if set for 5 second off) If in the scanning mode,
point at target, release the trigger as described.
6.
SETUP MODE
42 dB
25 kHz
002 Rec
S=60 A: 0dB
C: 0dB
B: 0dB
D: 0dB
Press ENTER
for Store (A)