5. USERS INSTRUCTIONS
TRISONIC
TM
SCANNING MODULE
:
1.
This module plugs into the front end of the instrument.
2.
Align the pins located at the rear of the module with the four jacks in the front end of the Metered Pistol Housing (MPH)
and plug in.
3.
For general use position the frequency selection to 40 kHz.
4.
Start to scan the test area.
METHOD OF AIRBORNE DETECTION:
The method of air borne detection is to go from the "gross to the fine". Start at maximum sensitivity (S=70), constantly reducing the
sensitivity and following the bar graph amplitude display to the loudest point. If there is too much ultrasound in the area, reduce the
sensitivity, place the RUBBER FOCUSING PROBE (described below) over the scanning module and proceed. Every time the sound
level rises to a point where it is difficult to follow, reduce the sensitivity again and again until you are able to follow the test sound to
its' loudest point.
HEADSET:
The DHC-HH headphones are designed to be worn with hard hats. To use, simply plug the headset cord into the headset Jack on the
pistol housing, and place the headphones over your ears.
RUBBER FOCUSING PROBE
:
To use, simply slip it over the front of the scanning module or the contact module.
NOTE: To prevent damage to the module plugs, always remove the module BEFORE attaching and/or
removing the Rubber Focusing Probe.
STETHOSCOPE MODULE:
1.
Align the pins located at the rear of the module with the four jacks in the front end of the Metered Pistol Housing (MPH)
and plug in.
2.
Touch test area.
As with the SCANNING MODULE, go from the "gross" to the "fine". Start at maximum sensitivity (S=70) on the Sensitivity
Selection dial and proceed to reduce the sensitivity until a satisfactory sound level is achieved. At times it may be necessary to utilize
the STETHOSCOPE MODULE with the sensitivity level at or near maximum. Occasionally when in this situation stray ultrasound
may interfere with clear reception and be confusing. If this occurs, place the RUBBER FOCUSING PROBE over the Stethoscope
probe to insulate against the stray ultrasound.
STETHOSCOPE EXTENSION KIT:
1.
Remove the Stethoscope Module from the Metered Pistol Housing.
2.
Unscrew the metal rod in the Stethoscope Module.
3.
Look at the thread of the rod you just unscrewed and locate a rod in the kit that has the same size thread - this is the "base piece".
4.
Screw the Base Piece into the Stethoscope Module.
5.
If all 31" (78.7 cm) are to be utilized, locate the middle piece. (This is the rod with a female fitting at one end) and screw this
piece into the base piece.
6.
Screw third "end piece" into middle piece.
7.
If a shorter length is desired, omit step 5 and screw "end piece" into "base piece".
TO CHARGE THE UP10000:
1.
Plug recharger cable into recharger jack on the UP10000 and then plug the recharger into a wall receptacle.
2.
Make sure that the LED on the charger is blinking when recharging.
3.
The LED remains solid when the battery is charged. The instrument may stay connected to the charger without
damaging the battery. Charge time is approximately 4 hours.
4.
WARNING
: Use the supplied UE Systems recharger (BCH-10L)
only
. Use of unauthorized rechargers will void the warranty
and may degrade or damage the battery.