background image

16

e.

RECORD WAV

: The Ultraprobe 10000 is configured to record heterodyned ultrasounds with this function setting.

f.

CF INFO

(Compact Flash): Information of stored recorded WAV files and record time can be viewed here.

g.

STORE REC

: Use this function to store inspection data. This is a quick way to data log if inspection data is to be saved

without viewing previously stored data (as in Storage Display).

Storage Display Information:

When in the

Storage Display

mode, the basic inspection parameter information can be viewed and stored if desired. To view

Parameter Information when in the

Storage Display

mode:

1.

“Click” the

Sensitivity Control Dial

until the “

Press ENTER For

” blinks.

2.

Spin to

Param INFO

and Press (Click) the

Yellow ENTER Button

to enter

3.

The information display will read: “

Test Results

:” and “

Temperature

:”

PARAMETER INFO (in Storage Display only)

The

Parameter Info

screen displays the test data relevant to a selected application (ex: bearings, valves, leak, steam). This

screen may be used to enter additional data such as test results or temperature. An extension of the

Parameter Information

screen is the

MORE

Function Selection. This will display additional inspection information: Date, Time, Module used and

Offset Value (if an offset value has been set). See the APPLICATIONS section for details. Note: this is the same information as
will be seen in the basic application information screens. i.e.:

GENERIC INFO

or

BEARING INFO

.

3.

ABCD NAVIGATION

Note: The

ABCD Display

will only work if the “

Valve

” application has been enabled in the

Setup Mode

. This is described in

SETUP, “Menu 03 Application Select

”. To select the

ABCD Screen

, after the

Valve

Application has been selected,

Select

Menu 02

,

Display Screens

in the

Setup Mode

and spin to

ABCD

.

Moving around the

Display Panel

is easy. Basically three clicks (presses) of the

Sensitivity Control Dial

will move the cursor

on the

Display Panel

to three key spots: the

Decibel Indicator

, the

Frequency Indicator

, and the

Function Selection Indicator

.

When an indicator is selected, or “active”, it will blink. When the cursor is moved to the

Decibel Indicator

and the

Decibel

Indicator

is blinking, it may be spun (rotated) clockwise or counter clockwise to adjust the sensitivity/volume of the instrument.

When the cursor is moved to the

Frequency Indicator

and the

Frequency Indicator

is blinking, the

Sensitivity Control Dial

may be spun to change the frequency. When blinking, the

Function Selection

, located on the bottom of the screen can be

changed by spinning the

Sensitivity Control Dial

up or down to select a specific feature which may be entered by pressing the

Yellow Enter Button

. Each A,B,C and D

dB

entry will be stored separately by clicking the

Yellow ENTER button

to store.

On the bottom line of the display panel, you will find the Function Select Indicator: “

Press Enter for:”

(you can use this to

enter any of the following default fields) The default function fields are:

a. STORE (A)

K. CF INFO (COMPACT FLASH)

b. STORE (B)

L. STORE RECORD AND EXIT

c. STORE (C)
d. STORE (D)
e. SETUP MENU
f. TEXT EDITOR (ON/OFF)
g. ALARM (DISABLED ENABLED)
h. VALVE INFO
i. STORAGE DISPLAY
j. RECORD WAV

Click for dB

Click for Frequency

Click for Function selection

Содержание 100-UP 10000SD

Страница 1: ...U L T R A P R O B E 1 0 0 0 0 I N S T R U C T I O N M A N U A L V 3 2 01...

Страница 2: ...Accessories 7 LRM 9 10 7 CFM 9 10 7 UWC 9 10 7 DHC 2 Headphones 7 TFSM 7 TFCM 7 UFMTG 1991 7 WTG 2SP Warble Pipe Threaded Tone Generator 7 BCH 10L 7 BCH WTG2 8 HTS 10 8 LLA 8 RAS RAM MT 8 3 Overview...

Страница 3: ...CD Information 19 Freezing a Reading on the Display Panel 19 7 Setup Mode Overview Of Setup Mode 20 How To Use The Setup Mode 20 Menu 01 Data Transfer 20 Menu 02 Display Screen 21 Menu 03 Application...

Страница 4: ...Level 34 36 Leaks UltratoneTM 37 Electric Arc Corona Tracking Detection 38 Monitoring Bearing Wear 39 Detecting Bearing Failure 40 Comparative and Historical Trending 40 Lack Of Lubrication 40 Over Lu...

Страница 5: ...en up a world of opportunities for inspection and data analysis ULTRASOUND TECHNOLOGY INSTRUCTION SCHOLARSHIP CERTIFICATE Your Ultraprobe 10000 has many applications ranging from leak detection and el...

Страница 6: ...amplification to allow for a clear intelligible signal to be received and interpreted For placement align the four prongs on the back with the corresponding receptacles in the front of the pistol and...

Страница 7: ...and the RAM Remote Access Module is plugged into the front end of the Ultraprobe B OPTIONAL ACCESSORIES CFM 9 10 A scanning module used for close proximity leak detection in pressure and vacuum system...

Страница 8: ...WTG2 This is an optional battery charger for use with all Warble Tone Generators The line input is 220 VAC 50Hz and the charging time is about 8 hours HTS 10 Holster set for the UP10000 Includes a ut...

Страница 9: ...out Offset see dB Scale Select in the Set Up Mode Section TRIGGER ON OFF TRIGGER SWITCH The Ultraprobe is always off until the trigger switch is pressed To operate press the trigger To turn the instru...

Страница 10: ...ayed in the display panel WRIST STRAP To protect the instrument against being dropped accidentally use the wrist strap SENSITIVITY CONTROL DIAL This is one of the most important controls in the unit I...

Страница 11: ...2 Touch test area As with the SCANNING MODULE go from the gross to the fine Start at maximum sensitivity S 70 on the Sensitivity Selection dial and proceed to reduce the sensitivity until a satisfacto...

Страница 12: ...he reading while holding the trigger in Immediately release the trigger and the reading will be set Once the trigger is pulled again the instrument will return to normal operation NOTE This operation...

Страница 13: ...wise or counter clockwise to adjust the sensitivity volume of the instrument When the cursor is moved to the Frequency Indicator and the Frequency Indicator is blinking the Sensitivity Dial may be spu...

Страница 14: ...w able to adjust the sensitivity 3 Once in the Sensitivity mode turn the Sensitivity Control Dial clockwise to increase the sensitivity and counter clockwise to decrease the sensitivity 4 The Sensitiv...

Страница 15: ...the Decibel Indicator the Frequency Indicator Storage Location Indicator and the Function Selection Indicator When an indicator is selected or active it will blink When the cursor is moved to the Dec...

Страница 16: ...the Valve application has been enabled in the Setup Mode This is described in SETUP Menu 03 Application Select To select the ABCD Screen after the Valve Application has been selected Select Menu 02 Di...

Страница 17: ...e bottom of the Display Panel until Press ENTER For is blinking Spin to Edit Text and press Enter 3 The text icon will blink If the field has no entry the screen will be blank 4 The Sensitivity Contro...

Страница 18: ...Temperature entry be desired 8 Click on the Temperature setting it will blink rapidly Spin to the left to select a 100 reading ex 200 300 and to the Right to select a tens reading ex 50 9 Click to set...

Страница 19: ...the information Click Enter a screen will pop up to confirm the information has been stored 3 To view previously stored information Click the Sensitivity Control Dial to the bottom of the Display Pane...

Страница 20: ...ent is off 1 Click both the Yellow ENTER button and the Sensitivity Control Dial at the same time Only after these two controls have been pressed then squeeze the trigger NOTE Hold the Trigger in duri...

Страница 21: ...ored data for a selected storage location 1 Follow the steps to Enter the Setup Mode as described earlier 2 Spin to Menu O2 Display Screens 3 Click the Sensitivity Control Dial to enter Display Screen...

Страница 22: ...blinking 2 Spin to Alarm Enable 3 Click to Enter 4 Spin to the desired dB Alarm level 6 Click the Yellow ENTER button to exit Menu 07 Set Record Time The recording time can be preset using this featu...

Страница 23: ...ext entry is not needed select OFF The Text Editor can be set by entering the Setup Mode 1 Spin to Text Editor 2 Click the Sensitivity Control Dial to enter the Text Editor enable mode 3 Spin the Sens...

Страница 24: ...s described earlier 2 Spin to Menu 13 Sensitivity Default 3 Click to Enter 4 Spin to selected sensitivity level 5 Click to set 6 Press the Yellow ENTER button to exit Menu 14 User Frequency Default Th...

Страница 25: ...Yellow ENTER button to exit Menu 18 Set Turnoff Time Turn off 0 3 minutes preset at 5 seconds 1 Follow the steps to Enter the Setup Mode as described earlier 2 Spin to Menu 18 Set Turnoff Time 3 Click...

Страница 26: ...ive sleeve negative Input level is 0 5 V DC maximum Input resistance is 50 k OHMS Transfer function 1 mV DC degree o C or o F Input range 10o to 475o o C or o F Menu 23 Sensitivity Mode MANUAL AUTOMAT...

Страница 27: ...nk rapidly 5 Spin to select a new value 6 Click to set 7 Spin to TIME setting and click on either Hour or Minute the displayed number will blink rapidly 8 Once an Hour or Minute has been selected spin...

Страница 28: ...el is sensed The bar graph will move up and down to display sound intensities but the Peak Hold dB reading in the upper left corner will remain constant A thin vertical line on the bar graph indicates...

Страница 29: ...e g battery discharges data will not be saved to the Compact Flash Card To select this feature 1 Follow the steps to Enter the Setup Mode as described earlier 2 Spin to Auto Store to CF Menu 33 blinks...

Страница 30: ...n such as date time module used and a command for data storage will be shown in the MORE function FIELD ENTRY LIST FOR GENERIC Test Results OK CHK TR2 TR9 B Valves NOTE To use the ABCD Display Screen...

Страница 31: ...ity Dial until the cursor moves to the bottom of the Display Panel until Press ENTER For is blinking Spin to Electric Info and press Enter 2 To enter a test condition 3 Spin to Test Results the condit...

Страница 32: ...e setting it will blink rapidly Spin to the left to select a 100 reading ex 200 300 and to the Right to select a tens reading ex 50 12 Click to set 13 Spin to Man Manufacturer 14 Click to enter and Sp...

Страница 33: ...re to enter and spin to desired pressure PSI or BAR 9 Click to Set 10 Spin to Application codes as APP0 APP9 11 Click to Enter and Spin to desired Application 12 Click to Set 13 Spin to Distance 14 Cl...

Страница 34: ...noise There are ultrasonic components in this white noise Since the ultrasound will be loudest by the leak site the detection of these signals is usually quite simple A leak can be in a pressurized s...

Страница 35: ...BE over the scanning module and proceed to scan the test area 9 Listen for a rushing sound while observing the meter 10 Follow the sound to the loudest point The Display will show a higher reading as...

Страница 36: ...he frequency until the weak signal appears to be clearer and then follow the basic detection methods previously outlined 2 LOW LEVEL LEAKS In ultrasonic inspection for leakage the amplitude of the sou...

Страница 37: ...Placement is dependent upon such variables as the size of the leak to be tested the thickness of the test wall and the type of material tested i e is it sound absorbent or sound reflective Remember yo...

Страница 38: ...ment follow all your plant or company safety procedures When in doubt ask your supervisor Never touch live electrical apparatus with the Ultraprobe or its accessories The method for detecting electric...

Страница 39: ...ny other indicators including heat and vibration changes An ultrasonic system based on detection and analysis of modulations of bearing resonance frequencies can provide subtle detection capability wh...

Страница 40: ...ality of the signal for proper interpretation 7 Select same type bearings under similar load conditions and same rotational speed 8 Compare differences of meter reading and sound quality PROCEDURE FOR...

Страница 41: ...end bearings including low speed bearings It can also extend the use of the FFT to record all types mechanical information such as leaking valves cavitation gear wear etc 3 GENERAL MECHANICAL TROUBLE...

Страница 42: ...r continuous flow operation of the trap Intermittent traps are usually the inverted bucket thermodynamic disc and thermostatic under light loads Continuous flow include the float float and thermostati...

Страница 43: ...e steam As the steam starts to condense the pressure against the disc lessens and the trap cycles A good disc trap should cycle hold discharge hold 4 10 times per minute When it fails it usually fails...

Страница 44: ...uations the ultrasound generated within the system will be so intense that surface waves will travel from other valves or parts of the system and make it difficult to diagnose valve leakage In this ca...

Страница 45: ...level decreases as you move towards the test valve it indicates that the valve is not leaking 6 If the sound level increases as you approach the test valve It is an indication of a leak in the valve...

Страница 46: ...be greatest at the site of the partial blockage PROCEDURE 1 Use Contact Stethoscope Module 2 Touch downstream side of suspected area and listen through headset 3 Start test at 40 kHz If the sound appe...

Страница 47: ...nd amplification is a very important feature Although there are obvious audible sounds emitted by most operating equipment it is the ultrasonic elements of the acoustic emissions that are generally th...

Страница 48: ...equency Battery Status and 16 Segment Bar Graph Sensitivity Detects 005 127 mm diameter leak 5psi 34 bar at a distance of 50 ft 15 24 m Threshold 1 x 10 2 std cc sec to 1 x 10 3 std cc sec Dimensions...

Страница 49: ...e case you will see a change lever Move this change lever to the middle of the lock so that it hooks behind the change notch picture 1 2 Now set your personal combination by turning the dials to the d...

Страница 50: ...50 14 Hayes Street Elmsford NY 10523 USA Tel 914 592 1220 Fax 347 2181 Email info uesystems com Web page http www uesystems com UP10M1009 2009 UE Systems Inc All rights reserved...

Отзывы: