38
Operative Instructions
2) Loosen and remove both spray arms.
3) Rinse the spray arms thoroughly.
4) Reassemble and tighten in position the spray arms.
a
WARNING! It is mandatory to record
the weekly spray arms cleaning in the
Maintenance Logbook. Daily cleaning is also
recommended to ensure continuous optimal
operation.
11.2.3 Cleaning the air suction filter
2
1
Fig. 26
1) Remove the filter cover panel (5 Fig. 2.1 and Fig.
2.2).
2) Unscrew the screws (1 and 2 Fig. 26) using a Torx
screwdriver T20.
3) Pull out the prefilter.
4) Clean the prefilter with compressed air or cold
water and dry the prefilter completely.
a
WARNING! Reassemble the prefilter only
once it is completely dry. If it is not completely
dry it can damage the device.
5) Reassemble the completely dry prefilter.
6) Retighten the screws and refit the filter cover panel.
a
WARNING! It is mandatory to record the
monthly air suction filter cleaning in the
Maintenance Logbook, to ensure continuous
optimal operation.
11.2.4 Replacing the air suction filter
The washer disinfector is standard equipped with a
suction filter for drying air.
The machine is also standard equipped with an
«absolute» «HEPA H14» class supplementary air filter,
in compliance with EN 1822.
a
WARNING! It is mandatory to replace
the prefilter and the HEPA H14 filter as
indicated in the Maintenance Logbook. In
case of intensive use of the machine it is
recommended to increase the frequency of
maintenance.
When replacing the air suction filter use a Torx
screwdriver T20 to open the filter panel.
11.3 Special maintenance
Special maintenance operations are not foreseen by
the user but must be performed exclusively by the
technical assistance or by an authorised and qualified
service personnel that must refers to the Service
Manual.
How to carry out special maintenance operations and
the frequency of maintenance are described in the
Service Manual.
Contact the Manufacturer service for special
maintenance
.
a
WARNING! Special maintenance is carried
out by the technician under guarantee only if
the uilizer has correctly and regularly carried
out all ordinary maintenance operations and
if the uilizer has filled out and kept up-to-
date the Maintenance Logbook.
11.4 Troubleshooting
The annex 12.8 lists the main malfunctions, with
relative causes and recommended solutions, that the
machine may present during its operation.
The interventions that may become necessary must be
carried out by experienced and qualified operators.
If the problems persist or appear even more often
after performing the work reported below, contact
the technical assistance or an authorised and qualified
service personnel.