background image

27

Operative Instructions

7)  Start

 

STD: enter the selected menu item; 

SPC: continue to the next parameter; 

SPC: "B" character for the custom password.

8)  Reset

 

STD: stop a running program; 

SPC: return to the previous menu item; 

SPC: "A" character for the custom password.

7.1  Colors of the keys

Depending  on  the  menu  function  you  are  using,  the 

keys get highlighted with different colors:

• White light: indicates the keys that can be used in that 

particular action.

• Red  light:  indicates  the  function  "go  back  to  the 

previous  menu"  if  associated  with  the  Door  key; 

indicates the function "delete the characters typed" 

if associated with the P+ key. 

• Green light: indicates confirmation of the action.

 

Ö

On  any  screen  and  for  any  action  you  can 

only use the highlighted keys.

7.2  Display

The  LCD  display  shows  the  status  of  the  machine. 

The images show the various phases and the current 

operation. For example: if the machine is loading water, 

the display will show the image of the tap, of the water 

flow meter and of the water being loaded. If water is 

cold, droplets will be blue and if water is hot, they will 

be red. The temperatures are set in degrees Celsius or 

Fahrenheit. The images are animated, to show how the 

current operation evolves.

Fig. 16

Description of the display (Fig. 16):

1) 

The  number  of  the  running  program  is  displayed 

(e.g. P02 = program 2);

2) 

It displays the phase the machine is running. If the 

machine is going to run a wash cycle it will display 

the  writing  «Wash».  When  the  machine  displays 

an alarm that space background turns red and the 

message shows the number of the alarm and a brief 

description;

3) 

Water  conductivity  value  in  microsiemens  (if 

conductivity probe function is enabled);

4) 

Several  animated  images  showing  the  current 

status of the machine are displayed;

5) 

The  elapsed  phase  time  from  the  moment  the 

temperature  set  for  the  phase  (see  12)  has  been 

reached is displayed;

6) 

Bar  indicating  the  progress  of  the  program;  if  the 

program  is  ending,  the  bar  will  be  almost  entirely 

green;

7) 

The  temperature  measured  by  the  PT1000  probe, 

placed after the air heater, is displayed to indicate 

the temperature of the air entering the chamber;

8) 

It  displays  the  temperature  measured  by  the  first 

PT1000 probe placed in the chamber;

9) 

It is the display of the temperature measured by the 

second  PT1000  probe  placed  in  the  chamber.  The 

temperature detectable by the two probes must not 

differ from each other by more than 36°F (2°C);

10) 

Time for which the device should maintain the set 

temperature (see 12);

11) 

Phase of the program;

12) 

Temperature set for the current phase;

13) The A

0

 value is displayed during disinfection;

14) 

Date  and  time  are  displayed  with  the  machine  in 

stand-by.

7.2.1  Messages on the display

On  the  display  of  rack  insertion,  messages  may  be 

displayed, such as:

• Product  1  liquid  reserve:  meaning  that  the  liquid 

inside  the  product  1  tank  is  finishing  and  must  be 

replaced;

• Product  2  liquid  reserve:  meaning  that  the  liquid 

inside  the  product  2  tank  is  finishing  and  must  be 

replaced.

When  the  cycle  has  finished  the  following  message 

appears:  «Finish»  and  a  green  light  appears  in  the 

chamber. At this point the door is unlocked and you can 

unload the rack with the washed items.

7.2.2  Bluetooth enabled 

If bluetooth is enabled, the following icon will appear 

on some display screens:

Содержание TIVA8-MD

Страница 1: ...perate with the washer disinfector read these instructions for use INSTRUCTIONS FOR USE Original instructions Code Man205 0745009EN Rev 02 Edit 06 2022 Language English Firmware version 00B03 WASHER D...

Страница 2: ...due to negligence and failure to comply with this manual and is also not liable for any damage caused by incorrect interpretations of the instructions contained Check for any updates to this manual o...

Страница 3: ...6 2 5 Treatment of ophthalmic instruments 24 6 3 Programs 24 6 3 1 Suggested programs vs configurations 24 6 3 2 Program structure 24 6 4 Starting the machine 24 6 4 1 Before starting the program 24 6...

Страница 4: ...r features The same personnel must store it and keep it intact to allow its consultation throughout the life span of the machine itself The personnel in charge must strictly and diligently follow the...

Страница 5: ...environment with a maximum humidity of 80 for temperatures up to 88 F 31 C decreasing linearly to 50 at 104 F 40 C 2 2 Label specifications Thewasherdisinfectorisequippedwithanidentification label for...

Страница 6: ...ance and assistance of the product contained in this use and maintenance manual have not been complied with If chemicals not identified by the manufacturer as compatible with the product are used With...

Страница 7: ...he power supply voltage on the line neutral 1200V and a duration up to 5s of long duration which may occur between the line conductor and the system earthing which may have a voltage equal to the powe...

Страница 8: ...oise during wash phases 56 dB A Max dBA noise during drying phase 61 dB A Type of protection according to IEC 60529 IP20 IP International Protection Against ingress of solid foreign objects 12 5 mm di...

Страница 9: ...level basic version 3b Washing rack with spray arm positioned at the upper level basic version 3c 1 1 DIN basket manufactured in stainless steel suitable for washing all kinds of instruments 3d Suppor...

Страница 10: ...zed people from using the machine The user undertakes to adequately inform and train its personnel on the application and observance of the safety regulations Failure to comply with these rules may je...

Страница 11: ...ER Do not wash the machine with direct or pressure water jets or corrosive substances Do not use the machine to wash objects and or containers that because of their shape or material are not compatibl...

Страница 12: ...operation takes place and check the integrity and suitability of the lifting and transportation equipment available Also make sure that the area of operation is clear and that there is sufficient esc...

Страница 13: ...ills 4 1 Storage If the packaged machine is temporarily stored make sure that it is not subjected to blows and tampering However it must be placed in a closed dry dust free environment and protected f...

Страница 14: ...supply in compliance with the current laws and regulations a WARNING In the single phase version the main switch of the power supply line must be a multi pole circuit breaker with adequate residual cu...

Страница 15: ...e disconnection procedure of the electrical connection by placing the dedicated safety device in OFF state The upstream electrical power line must be dimensioned and protected in accordance with curre...

Страница 16: ...IG 500 kPa 72 PSIG 500 kPa Connect the cold hot and deionized if available water hoses exiting the machine with their respective network connections as shown on the installation drawing It will be the...

Страница 17: ...ing the appropriate funnel a WARNING During this operation check that the plastic cap is closed 4 Insert the rack and start a normal washing cycle The machine regenerates automatically a WARNING The w...

Страница 18: ...ed manually regardless of the warning message on the display 5 4 Connection of the drain trap A B 1 1 3 2 2 Fig 10 Water and drain system Fig 10 A Floor drain connection B Wall drain connection 1 Hose...

Страница 19: ...ter 4 Salt tank cap Insert the supplied filters in the specific positions Constantly check the cleanliness of the filters especially the chamber bottom filter 3 Fig 11 Use this filter to have a high f...

Страница 20: ...ren and outsiders Possibly locked away Use on only the manufacturer s approved products Do not use liquids for household dishwashers 5 6 1 Refill procedure 1 Set up a new tank with the chemical 2 Remo...

Страница 21: ...be cleaned a WARNING Check that there are no water leaks in the pipe fittings 2 Check that the suction tubes located in the lower part of the machine within the corresponding tanks containing the liqu...

Страница 22: ...the door cannot be opened normally d HAZARD If the emergency release is used while a program is running very hot water and chemicals may leak Therefore there is a danger of burns scalding and irritat...

Страница 23: ...e baskets Before the automatic treatment check that the lumen and hollow body instruments can be treated internally and remove any obstructions Carefully place the load in the supports Objects must no...

Страница 24: ...ne leaves the factory with presetting washing programs already entered in the settings menu For programs details see the Washing Programs booklet 6 3 1 Suggested programs vs configurations For a good...

Страница 25: ...too high water temperatures Wash It is the actual cleaning cycle The cleaning chamber heats up until it reaches the specific temperature for the selected program a temperature that remains stable for...

Страница 26: ...burns a WARNING Failure to follow these recommendations can cause burns 7 Control Panel 1 2 3 4 8 7 6 5 Fig 15 The control panel consists of 8 keys and a 3 5 inch LCD display All keys except key 3 ar...

Страница 27: ...the number of the alarm and a brief description 3 Water conductivity value in microsiemens if conductivity probe function is enabled 4 Several animated images showing the current status of the machine...

Страница 28: ...and close the door WARNING The door must be tightly closed until you hear the classic closing click otherwise the program will not start Fig 18 1 2 With the machine started and door closed the screen...

Страница 29: ...t in the chamber Fig 18 6 Phase 2 6 Self loading of cold water During the self loading phase the chamber will be filled Fig 18 7 7 Prewash phase begins Fig 18 8 Phase 3 8 At the end of the prewash pha...

Страница 30: ...t Fig 18 12 Phase 5 12 At the end of the wash phase the water is automatically discharged Fig 18 13 Phase 6 13 Self loading of hot water During the self loading phase the chamber will be filled Fig 18...

Страница 31: ...of the water At the end of the phase if the conductivity will be below a presetted value the machine will procede to the next phase If the check is negative the machine will repeat phases 7 and 8 aga...

Страница 32: ...control panel and a red led light in the chamber At first the image concerning the alarm is displayed for 5 seconds then the alarm description is displayed for 10 seconds The image and text alternate...

Страница 33: ...consists of two screens the item when selected lights up Fig 20 3 Fig 20 4 7 2 7 Settings See Annexes 12 1 and 12 2 7 2 8 Language setup To change the display language with the door open press PRG fo...

Страница 34: ...this software in fact it s possible to create and print digital versions of the reports connect the device to the PC to customize the user names and passwords via USB port cable not included use USB...

Страница 35: ...ater The value increases by 1 C at a time ERR Alarms 0000 means that the machine is not in an alarm state When an alarm is triggered its code will be transmitted until the alarm is reset CYC Programs...

Страница 36: ...in this manual a WARNING The washer disinfector maintenance operations must be carried out with the machine completely switched off a WARNING If the machine is not used for more than 24 hours it is ne...

Страница 37: ...brushes bleach bleach cause the oxidation of the stainless steel surface and consequent change of color or any cleaning agent containing bleach or anything abrasive to cleaning the device Doing so wil...

Страница 38: ...WARNING It is mandatory to replace the prefilter and the HEPA H14 filter as indicated in the Maintenance Logbook In case of intensive use of the machine it is recommended to increase the frequency of...

Страница 39: ...are not disposed of correctly or are disposed of as household waste these components can damage human health and the environment Never dispose of the old washer disinfector among conventional waste U...

Страница 40: ...isables hot water If hot water is enabled purified water is automatically deactivated and vice versa OFF 8 User PIN enable at the start of cycle 0 OFF 1 ON Available only with User PIN Management opti...

Страница 41: ...Set password expiration time 6 22 User PIN enable at the end of cycle 0 OFF 1 ON Available only with User PIN Management option Enables or disables user PIN at the end of cycle OFF 23 Upper Spray Arm...

Страница 42: ...N Enables or disables the product 2 flowmeter ON 11 Product 3 enable 0 OFF 1 ON Enables or disables the product 3 pump ON 12 Product 3 flowmeter 0 OFF 1 ON Enables or disables the product 3 flowmeter...

Страница 43: ...program 6 Program cycles 40 Number of cycles performed with program 40 USER PIN 1 Pin User 1 To set the user password press the RESET key and enter the 6 desired numbers 2 Pin User 2 3 4 Pin User 20...

Страница 44: ...y language Resuming mode Program bu on 1 Program bu on 2 Program bu on 3 Printer language Drying enable Lower Spray Arm White LED always ON Fahrenheit temperature PIN expira on month User PIN enable a...

Страница 45: ...password change D O P S 3 Select User User Blocked Press Start to con rm P D P D P D P D P D P D Y N N Y N Enter old password Enter new password Con rm new password The password has been changed corr...

Страница 46: ...r user S P P 5 Select User Pin menu Insert Super User password max 3 a empts Digit password is valid is expired or rst access See password change procedure for Super User P D P D P D N N Y Y Select Us...

Страница 47: ...word expired Press Start to con rm N N Enter old password Enter new password Con rm new password The password has been changed correctly Is valid Is it equal to old password Y Y N Y S P P 5 Insert Sup...

Страница 48: ...WATER The hot water tap could be closed or partially open Check it Press RESET If it persists call the service personnel E22 NO DEMI WATER The demineralized water tap could be closed or partially open...

Страница 49: ...ight be broken or disconnected Press RESET If it persists call the service personnel E53 CHAMBER T DIFF A temperature difference between the probes in the chamber has been detected too high Press RESE...

Страница 50: ...D RESERVE PRODUCT 1 Product 1 is in reserve or may be empty Add the product 1 LIQUID RESERVE PRODUCT 2 Product 2 is in reserve or may be empty Add the product 2 LIQUID RESERVE PRODUCT 3 Product 3 is i...

Страница 51: ...s not reached The thermostat sensor in the washing chamber is covered with deposits Clean the thermostat sensor 5 The machine does not dry The air filter in the drying system is dirty or clogged Repla...

Страница 52: ...PT1000 2 temperature chamber control probe C Yes 0 3 1 0 31 0 17 18 19 20 TPT3 PT1000 3 temperature air probe C Yes 0 1 0 4 104 21 22 23 24 ERR Alarm number n Yes 0 0 7 0 70 25 26 27 28 29 30 CYC Coun...

Страница 53: ...0 0 6 6 71 72 73 74 CONP Conductivity probe status n Yes 0 0 0 2 2 75 76 77 78 79 A0 A0 Value n Yes 0 6 1 8 0 6180 80 81 82 83 SPRYR B Spray arm at the bottom of the chamber rotation monitoring rpm Y...

Страница 54: ...0 Program 40 PHA TYPE PHASE IN EXECUTION 0001 Drain 0010 Trolley insertion 0002 Prewash 0011 Door closure 0003 Wash 0012 Program selection 0004 Rinse 0013 Operator scan 0005 Disinfection 0014 Trolley...

Страница 55: ...150 150 ml of product delivered VCW VHW VDW QUANTITY OF WATER DELIVERED 0000 0 liters of water delivered liters of water delivered 0001 1 liter of water delivered 0012 12 liters of water delivered CON...

Страница 56: ...56 Operative Instructions Notes...

Страница 57: ...57 Operative Instructions Notes...

Страница 58: ...58 Operative Instructions Notes...

Страница 59: ...59 Operative Instructions...

Страница 60: ...Tuttnauer USA Co Ltd 25 Power Drive Hauppauge N Y 11788 USA Tel 1 631 737 48 50 1 800 624 58 36 Fax 1 631 737 07 20 Web www tuttnauerusa com...

Отзывы: