22
Operative Instructions
4) Wait a few seconds to see in the lower right part,
the area where the chemical inlet holes are present,
a constant flow free from air bubbles.
5) Release the button and repeat the sequence from
3 to 4 for the other dosing pumps that need to
recharge the hydraulic circuit.
1
2
3
4
8
7
6
5
Fig. 13
)
CAUTION! This phase must be performed
at the first installation of the machine, in
this case, let the liquid flow for a few more
seconds to ensure the correct and full
loading of the hydraulic system.
6.2 Before use
The washer disinfector can be used for cleaning and
disinfecting:
- medical and dental instruments;
- keys, trays, containers;
- hollow instruments, e.g. suction cannulas, fixing
them to the appropriate supports, using the suitable
adapters.
6.2.1 Emergency door release
In the event of a blackout, or any other need in which it
becomes difficult to open the machine door, there is a
manual emergency release that can only be activated if
the door cannot be opened normally.
d
HAZARD! If the emergency release is used
while a program is running, very hot water
and chemicals may leak. Therefore, there is a
danger of burns, scalding and irritation.
b
HAZARD! In the case of an emergency release
of the door, the load might be contaminated.
Be carefull while handling it.
Release the door:
1) Push the key supplied in the pack horizontally into
the gap between the door and the lid or worktop (1
Fig. 14.1), to position the key correctly, follow the
position indicated on the label (1 Fig. 14.2) on the
door handle.
2) Upon the door unlock take out the key.
1
Fig. 14.1
1
Fig. 14.2
In case of program interruption, proceed with a new
treatment of the load.
6.2.2 Preparing the load
a
WARNING! The maximum weight that the
open door can withstand is 66 lbs (30 kg).
The maximum volume of the wash chamber
is 44 US gal (165 l).
- The instruments to be washed must be placed on the
relative racks, making sure that they do not overlap
one another.
Ö
IMPORTANT NOTE! The cleaning quality
depends on the correct loading of the
instruments.
- Empty any residual liquids from the instruments and
containers before inserting them in the chamber,