48
Operative Instructions
ALARM
DESCRIPTION
ACTION
E33 FLOWMETER 4
Chemical liquid tank 4 might be empty.
Check it.
Press RESET
If it persists, call the service
personnel.
E34 PRODUCT TIMEOUT
The density of the chemical liquid might
be too high. Check it.
Press RESET
If it persists, call the service
personnel.
E41 DRAIN TIMEOUT
The drain pipe might be blocked or
clogged. Check it.
Press RESET
If it persists, call the service
personnel.
E43 DRY TEMPERATURE
The load might not be dried because the
minimum set temperature has not been
reached.
Press RESET
If it persists, call the service
personnel.
E44 PREWASH TEMPERAT
A too high temperature was detected
during the prewash phase.
Press RESET
If it persists, call the service
personnel.
E45 CHAMBER T.LIMIT
A too high temperature was detected
inside the chamber.
Press RESET
If it persists, call the service
personnel.
E46 AIR T.LIMIT
A too high air temperature was detected.
Press RESET
If it persists, call the service
personnel.
E47 PHASE T. LIMIT
A too high temperature was detected
during the current phase.
Press RESET
If it persists, call the service
personnel.
E50 PROBE 1 BROKEN
Chamber probe PT1000-1 (work) might
be broken or disconnected.
Press RESET
If it persists, call the service
personnel.
E51 PROBE 2 BROKEN
Chamber probe PT1000-2 (control)
might be broken or disconnected.
Press RESET
If it persists, call the service
personnel.
E52 AIR PROBE BROKEN
Air probe PT1000-3 might be broken or
disconnected.
Press RESET
If it persists, call the service
personnel.
E53 CHAMBER T. DIFF
A temperature difference between
the probes in the chamber has been
detected too high.
Press RESET
If it persists, call the service
personnel.
E66 NO HEATING
There is a problem with water heating.
Press RESET
If it persists, call the service
personnel.
E67 CONDENSER LEVEL
The water inside the steam condenser
has reached the maximum level allowed.
Press RESET
If it persists, call the service
personnel.