39
Operative Instructions
12 Annexes
12.1 Menu parameter table
ID Parameter Name
Min
Max
Description
Factory
Default
GENERAL
1
End cycle buzzer
0=OFF
1=ON
Turns the buzzer on or off at the end of
a cycle
ON
2
Warning alarm
buzzer
0=OFF
1=ON
Turns the buzzer on or off when an
alarm is triggered
ON
3
Buttons noise
0=OFF
1=ON
Turns the noise on or off when a button
is pressed
ON
4
Device 1
0= NONE
1= PRINTER
2= LAN
Enables or disables the printer (if
present)
NONE
5
Device 2
0= NONE
1= KEY USB
Enables or disables the data storage
feature on USB stick
KEY USB
6
Deionized water
enable
0=OFF
1=ON
Enables or disables deionized water. If
deionized water is enabled, hot water
is automatically deactivated, and vice-
versa
ON
7
Hot water enable
0=OFF
1=ON
Enables or disables hot water. If hot
water is enabled, purified water is
automatically deactivated, and vice-
versa
OFF
8
User PIN enable at
the start of cycle
0=OFF
1=ON
Available only with "User PIN
Management" option. Enables or
disables user PIN at the start of cycle
OFF
9
Autom. Door
unlock
0=OFF
1=ON
Enables or disables automatic door
unlock at the end of the cycle
OFF
10 Font Display
0=EUR-JAP
1=EUR-CYR
Sets the font map to European /
Japanese or European / Cyrillic
EUR-JAP
11 Display language
1
8
Selects the desired language for the
display
ENGLISH
12 Resuming mode
0= PHASE
RESTART
1= CYCLE
RESTART
2= STANDBY
Resuming mode after an alarm or
machine switch-off during the cycle. If
at 0, the program restarts from the last
phase in which the unscheduled stop
occurred. If at 1 the program restarts
each time from the beginning. If at
2 the program does not restart but
remains in standby
STANDBY
13 Program button 1
1
40
Selects the program to be combined
with button 1
1