18
Operative Instructions
A
B
1
1
3
2
2
Fig. 10
Water and drain system (Fig. 10):
A) Floor drain connection.
B) Wall drain connection.
1) Hose connection waste water.
2) Hose connection capacitor discharge machine.
3) Minimum connection point 3.9" (100 mm).
Maximum connection point 23.6" (600 mm).
Ö
IMPORTANT NOTE! The glasswasher drain
pipe must be maintained if backflow and
other drainage issues are to be avoided. Keep
the drain pipe free from kinks and knots, and
minimize the number of twists and turns in
the pipe. Be careful not to crush the pipe.
If the pipe has been crushed, kinked, or
otherwise damaged, consider replacing it to
prevent future drainage issues.
5.4.1 Drain connection
The drain connection must be properly installed
according to the applicable instructions.
b
HAZARD! If the waste connection is installed
incorrectly, the waste water could flow
backwards into the machine’s chamber.
5.4.2 Waste hoses
The hoses:
- must be laid out so that they are not bent, pinched
or tangled;
- may not be connected together before reaching the
connection point;
- may not hang below the machine's lower edge.
5.4.3 Connection point for waste
- The connection point must have a capacity of
9.3 GPM (35 l/min) and must have a diameter of at
least DN 1.5" (40 mm).
- The hoses may be hung up using the accompanying
hose holder.
Ö
IMPORTANT NOTE! The manufacturer is not
liable in case of environmental pollution due
to an incorrect use of the glasswasher.
5.5 Chamber filters and spray arms
5.5.1 Chamber filters
1
3
2
4
Fig. 11
Filter unit (Fig. 11):
1)
Central filter.
2)
Mesh filter.
3)
Chamber bottom filter.
4)
Salt tank cap.