If the machine is to be moved, it is advisable to keep the
packaging for any other relocation.
If this is not provided, dispose of the packaging
materials: cardboard, polystyrene and other, separated
by single material, sending them to the most
appropriate final destination, which may be recovery or
storage in landfills.
4.1 Storage
If the packaged machine is temporarily stored, make
sure that it is not subjected to blows and tampering.
However, it must be placed in a closed, dry, dust-free
environment and protected from atmospheric agents.
When restored, a careful preliminary examination of its
integrity by specialised personnel is necessary.
Absolutely avoid overlapping weights or foreign bodies
on the packaging and the machine.
5 Installation
It is advisable that only furniture for professional use
be positioned in the area around the machine, to avoid
ruining them due to possible leakage of condensation
water.
b
HAZARD: Do not install and/or use the
thermodisinfector in environments with
flammable/explosive atmosphere.
a
Make sure that the floor is fit to support the
load of the equipment when in operation,
equal to 154,3 lbs. (70 kg). Make sure that
the machine is perfectly vertical and stable,
using a spirit level, if necessary.
Unevenness of the machine surface and height can be
adjusted with the four adjustable feet placed under the
machine itself.
There are two models of thermodisinfector (see Fig. 2):
The CT model is the classic model of thermodisinfector
with small-sized product bowls, which can be positioned
on any piece of furniture capable of withstanding its
weight, within a dental practice.
The UC model is a thermodisinfector with much more
capable product bowls than the TIVA2-C version and
which, usually, is embedded under a piece of furniture
where appropriate space is already provided, within a
dental practice.
b
HAZARD. All electrical and water connections
(loading/unloading) can only be carried out
by specialised and authorised personnel and
by consulting the relative diagrams.
Before machine positioning make sure that:
All the components needed for installation and proper
use of the machine were installed: main switch, water
supply taps, drain and anything else needed. These
components should have all the needed features and
should be installed at the locations shown on the
installation scheme.
a
WARNING: The use of unsuitable parts,
and/or the implementation of installation
procedures other than those shown on the
installation diagram, will immediately void
the machine warranty.
The characteristics of the electricity network must
be compatible with the values required for correct
operation indicated on the machine identification plate
and on the technical data sheet.
The machine must be connected to an efficient earthing
system (according to electrical safety standards).
Ö
The manufacturer is not to be held liable for
any damage caused by improper earthing of
the machine or faulty power supply.
0.0
9
”(
2
,5m
m)
0.0
9
”(
2
,5m
m)
1
.
9”
(
50
mm
)
CT = 25.6”(650mm) UC = 32.3”(820mm)
Fig. 7