Page 99
Message /
Symbol /
Problem
Failure Description
Corrective Action
than 1 second below the
sterilization temperature during
sterilization cycle.
High Temp
This message is displayed if
the temperature raises 7°F
(4°C) above sterilization
temperature during the
sterilization stage for 2
seconds during sterilization
cycle.
Perform a new cycle.
High Temp.
(Ending)
This message is displayed if
the system cannot reach the
required temperature, in the
chamber, within 10 minutes.
Perform a new cycle.
Heat Time
Error
This message is displayed if
the system cannot reach the
required temperature, in the
chamber, within the preset
time.
Verify that the autoclave
is not overloaded.
Heat Time
Error
(Keep)
This message is displayed if
the system cannot reach the
required temperature, in the
chamber, during the optional
"Keep Heat" stage, within the
preset time.
Verify that the autoclave
is not overloaded.
Low
Pressure
This message is displayed if
Chamber Pressure drops
below the sterilization pressure
(134°C = 304 kPa, 121°C =
205 kPa) for 2 seconds during
the sterilization stage.
Perform a new cycle.
High
Pressure
This message is displayed if
Chamber Pressure raises 4.2
psi-29 kPa above sterilization
pressure (134°C = 304 kPa,
121°C = 205 kPa) for 2
seconds during the sterilization
stage.
Perform a new cycle.
High
Pressure
(Ending)
This message is displayed if
the system cannot reach
atmospheric pressure ± 5kPa
Perform a new cycle.