Page 44
cycle. To remove such water, press Start and repeat the cycle. If the
problem persists, call for service.
•
(PV, pre-vacuum models only)
Drying
: vacuum is built up and the air
pump is working.
•
Ending:
The system checks that End temperature and End pressure
have been reached, and then releases the door.
5.9
Program 9: Bio Hazard Liquids (BH, bio-hazard
models only)
All exhaust from the chamber before completion of the
Sterilization stage is performed through the bio-hazard filter
. For
liquids when the manufacturer recommends autoclaving at
temperatures of 121ºC (250
F) for 45 minutes.
Cautions!
Both PT100 temperature sensors must be inside the bottles.
For proper sterilization, fill the bottles with approximately the
same amount of liquid.
Nominal parameters default settings
•
Sterilization temperature: 121
C (250
F)
•
Sterilization time: 45 minutes
Operations sequence:
•
(PV, pre-vacuum models only)
Pulse low/Pulse high:
at one pulse it
will build vacuum down to 25 kPa.
All exhaust from the chamber is
performed through the bio-hazard filter.
•
Water Inlet
: Water enters the chamber