Page 47
•
Cooling:
N/A.
•
Fast Exhaust:
the steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
•
Drying:
N/A.
Note:
Some water may remain in the chamber.
5.12
Program 12: Isothermal
For liquids when the manufacturer recommends autoclaving at 60-
100
C (140-212
F) with no drying. Recommending for melting of agar,
pasteurization etc.
Caution!
This is not a sterilization program!
Put one PT100 inside the bottle, leave the second one hanging
in the chamber outside the bottle (see below).
Nominal parameters default settings
•
Heating temperature (default): 80
C (176
F).