Page 13
On all models, a printer is an optional addition to the autoclave. The
printer prints the preset and actual parameters of the cycle
(temperature, time and pressure/vacuum).
The autoclave is provided with a pressure gauge that is used as guide
only. Should there be a power failure during the operation of the
autoclave, the pressure gauge indicates to the operator that there is
pressure in the chamber.
A deviation of +1.6% is accepted.
After operating the sterilizer, brown stains might appear on the bottom
of the chamber. These stains are a result of the heating elements that
are located at the lower external part of the chamber. The brown color
is a common phenomenon, can easily be removed, and will not have
any effect on the sterilized goods.
This manual is intended to give the user a general understanding of
how the autoclave works and indicates best ways to operate and
maintain it in order to obtain optimum results and a trouble free
operation.
After reading this manual, operating the autoclave should be straight
forward. However, since the autoclave is built using high technology
sensitive components, no attempt should be made by the user or any
other unauthorized person to repair or re-calibrate it.
Caution!
Only technical personnel having proper qualifications, holding
technical documentation and adequate test instrumentation, are
authorized to undertake repair or service.
3.2
Electrical Data
230V configuration
Property
Value: LABSCI 15 Value: LABSCI 15+L
Total Power
3000W (2*1500W)
4500W (3*1500W)
Voltage
1Ph, 230VAC
3Ph, 230VAC
Amperage
13A
13A
Protection against electrical
shock
Class I (IEC 60601-1)
Mains supply fluctuation
+/- 10%
Degree of protection by
enclosure
IP31