Page 21
3.10
Construction
The main parts of the autoclave are made of materials as indicated
below:
•
Chamber and door are made of stainless steel
•
Trays are made of stainless steel.
•
Water reservoir is made of hard plastic material.
•
Door handle is made of hard plastic material, which is safe to touch
and thermo-insulated.
3.11
Environment Emission Information
•
The peak sound level generated by the autoclave is less than 70 dBa
with background noise of 60 dBa.
•
The total heat per hour transmitted by the autoclave3840, 3850 models
is < 300 W/h, and for 3870 model is < 450Wh.
3.12
Safety Features
This autoclave includes built-in safety features such as:
•
Error message display.
•
Temperature dependent door locking system according to European
standards.
•
Electronic pressure and temperature measurement.
•
Safety relief valve to avoid build-up of excessive pressure.
•
Door switch enabling operation to be started only when the door is
closed.
•
Water level safety device.
•
Excess temperature protection.
3.13
Stickers Description
Symbol
Meaning
Part Number
Location
Caution! Hot
steam.
LAB048-0058
Near the safety
valve
Protective earth
(Ground)
LAB048-0020
Near the earthing
screw