Page 89
12.5.2
PED-approved type safety valve
1.
Operate the sterilization cycle according to the manual.
2.
Allow a pressure of approx. 200 kPa (29-psi) to build up in the
chamber.
3.
Remove the water reservoir cover.
4.
Turn the pressure relief nut counterclockwise for 2 seconds. Verify
steam escapes from the valve.
Attention:
Use protective gloves in order not to burn your hands with the hot
steam.
5.
Press the STOP key to interrupt the operation, and exhaust steam from
the chamber.
6.
Wait until the pressure decreases to zero, only then the door can be
opened.
12.6 Cleaning the water sensor
It is required that the water sensor be cleaned at least once per week.
Cleaning the sensor will ensure that the water level in the chamber is
properly reported to the microprocessor all during the cycle.
The water sensor is located in the rear of the chamber. It is easily
cleaned using a damp cloth or sponge, you may use a mild soapy
solution if you like. It is important to wipe the sides of the sensor as well
as the tip, to remove any dirt or debris that may have built up.
2