Page 37
All exhaust from the chamber before completion of the
Sterilization stage is performed through the bio-hazard filter.
For
instruments, when the manufacturer recommends autoclaving at
temperatures of 134
C (273
F) for 30 minutes. Drying stage is
available for PV (pre-vacuum) models only.
Nominal parameters default settings
Sterilization temperature: 134
C (273
F)
Sterilization time: 30 minutes
Drying time: 1 minute. (PV – pre-vacuum models only).
Operations sequence:
(PV, pre-vacuum models only)
Pulse low/Pulse high:
at one pulse it
will build vacuum down to 25 kPa.
All exhaust from the chamber is
performed through the bio-hazard filter.
Water Inlet:
Water enters the chamber
Heating:
The chamber is heated by actuation of electrical heaters
until the sterilization temperature is reached.
Sterilization:
Sterilization temperature is maintained constant during
the sterilization time.
Cooling:
N/A
Fast Exhaust:
the steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
Note:
If the cycle fails, fast exhaust is performed through the bio-hazard
filter. Some water may remain in the chamber after a failed cycle. To
remove such water, press Start and repeat the cycle. If the problem
persists, call for service.
(PV, pre-vacuum models only)
Drying:
vacuum is built up and the
air pump is working.
Ending:
The system checks that End temperature and End pressure
have been reached, and then releases the door.