400V configuration
Property
Value: 3840 Value: 3850 Value: 3870
Total Power
300W
(2*1500W)
3000W
(2*1500W)
4500W
(3*1500W)
Voltage
3Ph,
400VAC
3Ph,
400VAC
3Ph,
400VAC
Amperage
13A
13A
7A
Protection against electrical
shock
Class I (IEC 60601-1)
Mains supply fluctuation
+/- 10%
Degree of protection by
enclosure
IP31
3.3 Storage Conditions
Caution!
Packed or unpacked, the autoclave shall be retained in indoor
conditions!
3.4 Operating Conditions
The autoclave is intended to work in ‘indoor’ conditions only.
Only autoclavable materials shall be used.
The environment shall not exceed an ambient temperature range of
5
º
C (41
º
F)-40
º
C (104
º
F) and a relative humidity of 85% respectively.
The operation altitude shall not be over 2000 meters (6561 feet)
(ambient pressure shall not be lower than 80 kPa (11.6 psi)).
Cautions!
Waste water should be brought into the public net in accordance
with the local rules or requirements i.e. ONLY NON-HAZARDOUS
LIQUIDS SHALL BE DISPOSED IN PUBLIC SEWAGE!
The autoclave shall not be used in a manner not specified in this
manual! If the equipment is used in a manner not specified by the
manufacturer, the protection provided by the equipment may be
impaired.