3.9 Utilities
3.9.1
Electrical utility
Property
Value: 3850/3840
Value: 3870
Power supply
208/230VAC ±10%, 50Hz 400VAC ±10%, 50Hz
Recommended circuit
breaker
15A
10A
3.9.2
Other Utilities
Compressed Air (C
models only)
1/2" 3 Bar (44-58 psi)
Tap water
1/2", 2-6 Bar (29-44 psi)
Mineral free water
1/2", 2-3 Bar (29-44 psi)
Drain
2" Minimum Withstanding temp. of 80°C
(176°F)
Attention:
A switch or circuit-breaker must be included in the building installation.
This switch or circuit-breaker shall be in close proximity to the
equipment, within easy reach of the operator; and marked as the
disconnecting device for the equipment.
The electrical net must be protected with a current leakage safety
relay.
The electrical network must comply with local rules or regulations.
Verify that there is an easy access to the main power switch, to the
water cut-off valve and to the current leakage safety relay.
Make sure while placing the autoclave, to leave space around the
machine, to give the technician access to service the machine.
All water connections to autoclave must be performed through "BACK
FLOW PREVENTION SYSTEM" only, as per IEC 61770.