Page 94
Message /
Symbol /
Problem
Failure Description
Corrective Action
during the ending stage.
High
Pressure
(Exhaust)
This message is displayed if
the system cannot reach
preset pressure within 10
minutes from the beginning of
the exhaust stage.
Perform a new cycle.
Pressure
Time Error
This message is displayed if
the
system cannot reach the
required pressure conditions in
the chamber, after preset time,
during the air removal stage.
Verify that the autoclave
is not overloaded.
RTC Error -
Please Set
Current
Date and
Time
This message is displayed in
order to set the date and the
time.
Set Current Date And
Time.
If the problem persists,
call the technician.
Time Error
This message is displayed if
the real time clock is faulty.
Call the technician.
Door is
open
(During the
cycle)
This message is displayed
when the door is open: During
the cycle.
Close the door to
perform a new cycle.
Canceled
By User
This message is displayed
after the START/STOP key is
pressed and cycle aborted.
Wait until "cycle failed –
canceled by user" or
"cycle end – canceled by
user" is displayed.
Perform a new cycle.
Cycle
Failed
This message and symbol are
displayed if an error occurs
before sterilization cycle is
completed.
Perform a new cycle.
Air Error
This message is displayed at
the end of the cycle If the
autoclave does not reach the
atmospheric pressure after 10
minutes.
Wait until the autoclave
reaches the atmospheric
pressure and perform a
new cycle.
Compresse
d air supply
error
This message is displayed in
case of a compressed air
supply malfunction.
Check and fix the
compressed air supply.
Periodical
The periodical maintenance
Call for service.