Page 95
Message /
Symbol /
Problem
Failure Description
Corrective Action
check time
exceeded -
Please call
for service
time has passed.
Cycle
counter
exceeded -
Please call
for service
Number of cycles, since last
periodical maintenance,
exceeded the preset number
as defined by "cycle counter"
parameter.
Call for service.
Power
Down
This message is displayed if
power down has occurred
during the cycle. (this
message will print out in the
printer after the autoclave will
turn on).
Turn on the autoclave
and wait until the
autoclave is ready
(reaches the safe
condition) and perform a
new cycle.
Supply
distilled
water error
(digitat
input
option)
This message is displayed in
case of a mineral free water
supply malfunction.
Check and fix the
mineral free water
supply
Supply
water error
(digitat
input
option)
This message is displayed in
case of a city (tap) water
supply malfunction.
Check and fix the city
(tap) water supply.
Compresse
d air supply
error
(digitat
input
option)
This message is displayed in
case of a compressed air
supply malfunction.
Check and fix the air
supply.
No Water
This message is displayed if
the electrode in the chamber
did not sense water within the
preset time.
1.
check and fix the
mineral free water
supply.
2.check and clean the
water inlet filter.
3.
Clean the water
level electrode.