Wear heat-resistant gloves or use the tray handle to remove the load
from the autoclave
•
On completion of the cycle, the load shall be visual inspected to
ascertain that the load is dry, and that the colour of the sterilization
indicators turned to the required colour.
10.8
Stopping the process and cancelling the ERROR
message
•
It is possible to stop the program while the autoclave is operating.
Pressing the
START/STOP
key at any stage of the process stops the
operation. If the cycle was aborted before completing the sterilization
stage, it will leave the load unsterilized.
•
At the end of the aborted process (before completing the sterilization
stage),
"Cycle Failed"
message, error message and a warning
symbol
are displayed on the screen. Refer to "Displayed Error
Messages/Symbols".
•
Pressing the
START/STOP
key cancels the displayed message and
enables opening the door.
•
•If the door is not opened, the vacuum pump will stop automatically
after 30 minutes (ELPV model only).
Warning
The load has not completed a sterilization cycle, therefore it is
not sterile. Handle it as contaminated load.
10.9
Start Cycle by Clock mode
•
This mode enables the operator to define the time of the beginning of
the cycle. The maximum possible delay is 24 hours. For further
information see sec. 6.8.1.
“Start Cycle by Clock”.
10.10
Moving the Autoclave
1.
Disconnect the power supply cord.
2.
Disconnect the water and drain hoses.
3.
Disconnect the compressed air hoses (if applicable).
4.
Drain the water from the chamber.
To avoid injuries, Moving the autoclave should be done by using a
forklift.